What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318
- Lot or code
- Lot AZ4824290
- FDA recall number
- Z-1237-2026
- Quantity
- 427
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Tornier, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.