Tornier, Inc recalls
3 notices naming Tornier, Inc as the recalling firm, since April 30, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Tornier, Inc: Tornier HRS Max parts: 1. TORNIER HRS Max Tuberosity Body Lat Coat
Hazard (FDA): Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.
FDAMedical devices
Tornier, Inc: TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE
Hazard (FDA): A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected…
FDAMedical devices
Tornier, Inc: stryker Latitude EV HUMERAL REVISION STEM TRIAL
Hazard (FDA): A specific lot of Latitude Humeral Trial Stems has the incorrect direct part marking. " Laser Marked: 9030103, Humeral Stem Trial 150mm Large / Right, Yellow color indicator (incorrect) " Device Actually: 9030092…
FDAMedical devices