Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 97919: Philips North America Llc: Incisive CT and 1 more

FDA recall 97919 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • Incisive CT; Software Version Number: 4.5, 5.0, 5.1;

    Lot or code
    1) Model Number: 728143; UDI-DI: 00884838085015; Serial Numbers: 500092, 500202, 500309, 500376, 500574, 500265, 500264, 510051, 500034, 500544, 500543, 500548, 500046, 500168, 500064, 500110, 500111, 500102, 500145, 500178, 500255, 500217, 500211, 500216, 500225, 33009, 500242, 500252, 500222, 500282, 33005, 500301, 33001, 33035, 33039, 33002, 33003, 33007, 33013, 33047, 33011, 33008, 33015, 33020, 33010, 33046, 33017, 33019, 33049, 33016, 33024, 33006, 33025, 33034, 33041, 33038, 33031, 33042, 33033, 33037, 33045, 33012, 500032, 500073, 500138, 500308, 33023, 500443, 500410, 500412, 500557, 500038, 500069, 500332, 500397, 510056, 500091, 500135, 510028, 510029, 500591, 500123, 500087, 500136, 500103, 500116, 500105, 500056, 500125, 500041, 500089, 500200, 500152, 500219, 500256, 500213, 500307, 510001, 500233, 500371, 500370, 500377, 510030, 510032, 510031, 500360, 510005, 510020, 500329, 500361, 500436, 500457, 500406, 500465, 500525, 500535, 510027, 500523, 33043, 500028, 500024, 500122, 500173, 510095, 500374, 500003, 500020, 500052, 500075, 500118, 500149, 500150, 500134, 500146, 500218, 500223, 500232, 500299, 500311, 500394, 500435, 500463, 33028, 500480, 500078, 500157, 500153, 500229, 500276, 500310, 500300, 500318, 500336, 500347, 500423, 500372, 500357, 500597, 500351, 500363, 500414, 500390, 500448, 500598, 500399, 500405, 500428, 500391, 500524, 500498, 500422, 500526, 500542, 500541, 500565, 500575, 500578, 500592, 500585, 500593, 500586, 500604, 500581,…
    FDA recall number
    Z-0881-2026
    Quantity
    2341 units

    The text above is the beginning of the record; 17,281 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • CT 5300; Software Version Number: 4.5, 5.0, 5.1;

    Lot or code
    1) Model Number: 728285; UDI-DI: 00884838113237; Serial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;
    FDA recall number
    Z-0882-2026
    Quantity
    6 units (OUS only)

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

Source

Issued by
U.S. Food and Drug Administration
Firm
Philips North America Llc (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Philips North America Llc, Month: December 2025, Week 2025-W50. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97919 on December 10, 2025.

Other notices of Philips North America Llc