Philips North America Llc recalls
22 notices naming Philips North America Llc as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Crash
Philips North America Llc: Philips Medical Systems Technologies Ltd. Spectral CT 7500
Hazard (FDA): The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional…
FDAMedical devices
Philips North America Llc: Achieva 1.5T and 11 more
Hazard (FDA): A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
FDAMedical devices
Philips North America Llc: Philips Brilliance. Models: Brilliance iCT and 3 more
Hazard (FDA): Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
FDAMedical devices
Philips North America Llc: Cardiac Workstation 5000. Product Number: 860439. and 1 more
Hazard (FDA): Device will intermittently provide inaccurate data related to native heart rate.
FDAMedical devices
Philips North America Llc: Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following…
Hazard (FDA): Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
FDAMedical devices
Philips North America Llc: Philips Avalon Fetal Monitor and 1 more
Hazard (FDA): Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
FDAMedical devices
Philips North America Llc: DigitalDiagnost C90 Flex/Value/Chest/ER. Model Number: 712035.
Hazard (FDA): System was delivered without the required Seismic kit for Wall Stand VS2, which is designed to prevent damage and ensure stability of the Wall Stand VS2 during seismic events.
FDAMedical devices
Philips North America Llc: Brilliance iCT and 3 more
Hazard (FDA): As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners…
FDAMedical devices
Philips North America Llc: Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.
Hazard (FDA): It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.
FDAMedical devices
Philips North America Llc: Incisive CT and 1 more
Hazard (FDA): If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems…
FDAMedical devices
Philips North America Llc: IntelliVue MP2. Product Number: M8102A. and 21 more
Hazard (FDA): Potential issue where the IntelliVue monitors did not alarm.
FDAMedical devices
Philips North America Llc: Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel and 9 more
Hazard (FDA): The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
FDAMedical devices
Philips North America Llc: Spectral CT. Computed tomography X-ray system.
Hazard (FDA): Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT…
FDAMedical devices
Philips North America Llc: Product Name: SmartPath to dStream for 1.5T and 17 more
Hazard (FDA): Potential for alignment errors in the cross reference line functionality when reviewing images generated with the MobiView application.
FDAMedical devices
Philips North America Llc: Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub…
Hazard (FDA): Disconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.
FDAMedical devices
Philips North America Llc: Philips Brilliance CT Big Bore
Hazard (FDA): Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
FDAMedical devices
Philips North America Llc: Zenition 50. Product Code 718096. The devices are used for radiological guidance and… and 2 more
Hazard (FDA): A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
FDAMedical devices
Philips North America Llc: Philips DXR System and 1 more
Hazard (FDA): Potential for collimator to fall as a result of incorrect installation.
FDAMedical devices
Philips North America Llc: IntelliBridge EC10 Module. Model Number: 865115 and 4 more
Hazard (FDA): Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
FDAMedical devices
Philips North America Llc: Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Hazard (FDA): The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to…
FDAMedical devices
Philips North America Llc: Spectral CT and Spectral CT Plus. Model Numbers: 728333
Hazard (FDA): Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID…
FDAMedical devices
Philips North America Llc: Achieva TX Interventional Coil 3.0T and 12 more
Hazard (FDA): Potential safety issue where a patient may be harmed while preparing for or during a scan.
FDAMedical devices