What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
IntelePACS - InteleConnect / TechPortal
- Lot or code
- UDI-DI: B228INTELEPACS0. Version/Dates of Distribution: PACS-5-6-1-R64 through R68/02 Jul 2025 - 10 Sep 2025, PACS-5-7-1-R42 through R47/26 June 2025 - 25 Sep 2025, PACS 5-8-1-R20 through R23/26 Jun 2025 - 25 Sep 2025, PACS-5-9-1-R1 through R3/30 Jun 2025 - 09 Sep 2025
- FDA recall number
- Z-0723-2026
- Quantity
- 19
Distribution (FDA): US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- INTELERAD MEDICAL SYSTEMS INCORPORATED (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.