What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
IntelePACS (Image Fusion Module) - InteleViewer
- Lot or code
- UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later
- FDA recall number
- Z-1208-2025
- Quantity
- 88
Distribution (FDA): Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- INTELERAD MEDICAL SYSTEMS INCORPORATED (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.