What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
- Lot or code
- UDI-DI: 07640135140627, 07640135140603, 05415067036414 IFU: Part(Name) Revision: For US: 1. ARTEN600254920 (IFU CentriMag Blood Pump ABT) Rev C 2. ARTEN600254958 (IFU PediMag Blood Pump US ABT) Rev C 3. ARTEN100174794 (CentriMag Adult Pre-connected Pack IFU, US) Rev B 4. ARTEN600230157 (IFU CentriMag Blood Pump ECMO US) Rev D FOR OUS: ARTEN600254935 (IFU CentriMag Blood Pump CAN ABT) Rev C ARTEN600254936 (IFU PediVas Blood Pump Canada ABT) Rev C ARTEN600332205 (IFU CMag VAS-ECMO Pump OUS) Rev A ARTEN600332207 (IFU PVAS VAS-ECMO Pump OUS) Rev A REF 201-20003, 104318, 201-51006, 201-51007, 201-90001 201-90001T don't have associated un-expired batches/serial numbers. REF/Batch(Serial): 102953/10007334, 10007335, 10007336, 10012012, 10012013, 10012014, 10012015, 10012016, 10036179, 10036180, 10036181, 10037990, 10037992, 10037993, 10037995, 10037996, 10046028, 10046029, 10046030, 10046031, 10046032, 10046033, 10058754, 10058755, 10058756, 10061673, 10061675, 10061676, 10061677, 10065946, 10065947, 10065954, 10065955, 10065956, 10071358, 10071359, 10071361, 10073780, 10073781, 10073783, 10082053, 10082055, 10103208, 10103209, 10106345, 10106346, 10106347, 10106348, 10120074, 10120075, 10120076, 10120077, 10122157, 10122158, 10122159, 10125562, 10125563, 10125564, 10125565, 10138336, 10138337, 10138338, 10138339, 10145124, 10145125, 10145126, 10145127, 10164040, 10164041, 10164044, 10168193, 10168194, 10168195, 10168196, 10168197, 10168199, 10168200, 10168201, 10168202, 10168203,…
- FDA recall number
- Z-1099-2026
- Quantity
- 8399
The text above is the beginning of the record; 21,346 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Thoratec LLC
- Title