What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
392 more codes are in the record below.
HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.
- Lot or code
- UDI-DI: 05415067038234. MPUs distributed between April 2024 and February 2025. Serial Numbers: 20226263, 20226264, 20226265, 20226266, 20226267, 20226268, 20226269, 20226270, 20226271, 20226272, 20226273, 20226274, 20226275, 20226276, 20226277, 20226278, 20226279, 20226280, 20226281, 20226282, 20226283, 20226284, 20226285, 20226286, 20226287, 20226288, 20226289, 20226290, 20226291, 20226292, 20240622, 20240623, 20240624, 20240625, 20240626, 20240627, 20240628, 20240629, 20240630, 20240631, 20240632, 20240633, 20240634, 20240635, 20240636, 20240637, 20240638, 20240639, 20240640, 20240641, 20240642, 20240643, 20240644, 20240645, 20240646, 20240647, 20240648, 20240649, 20240650, 20240651, 20240652, 20240653, 20240654, 20240655, 20240656, 20240657, 20240658, 20240659, 20240660, 20240661, 20240662, 20240663, 20240664, 20240665, 20240666, 20240667, 20240668, 20240669, 20240670, 20240671, 20240672, 20240673, 20240674, 20240675, 20240676, 20240677, 20240678, 20240679, 20240680, 20240681, 20240682, 20240683, 20240684, 20240685, 20240686, 20240687, 20240688, 20240689, 20240690, 20240691, 20254123, 20254124, 20254125, 20254126, 20254127, 20254128, 20254129, 20254130, 20254131, 20254132, 20254133, 20254134, 20254135, 20254136, 20254137, 20254138, 20254139, 20254140, 20254141, 20254142, 20254143, 20254144, 20254145, 20254146, 20254147, 20254148, 20254149, 20254150, 20254151, 20254152, 20254153, 20254154, 20254155, 20254156, 20254157, 20254158, 20254159, 20254160, 20254161, 20254162,…
- FDA recall number
- Z-1570-2025
- Quantity
- 1661
The text above is the beginning of the record; 15,209 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Distribution (FDA): Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS, AL, CO, ME and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CZECH REPUBLIC, FINLAND, FRANCE, GEORGIA, GERMANY, HONG KONG, ISRAEL, ITALY, JAPAN, KUWAIT, LEBANON, LITHUANIA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, PORTUGAL, ROMANIA, SERBIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UK, UNITED ARAB EMIRATES.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Thoratec LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.