Skip to content
FDAMedical devicesClass IOngoing

FDA recall 96471: Thoratec LLC: HeartMate Mobile Power Unit: REF: 107754

FDA recall 96471 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

392 more codes are in the record below.

  • HeartMate Mobile Power Unit: REF: 107754, NA; 107758, EU; 107758AU, AU; 107758UK, UK; L107758, EU Rental. Used with HeartMate 3 and II Left Ventricular Assist System.

    Lot or code
    UDI-DI: 05415067038234. MPUs distributed between April 2024 and February 2025. Serial Numbers: 20226263, 20226264, 20226265, 20226266, 20226267, 20226268, 20226269, 20226270, 20226271, 20226272, 20226273, 20226274, 20226275, 20226276, 20226277, 20226278, 20226279, 20226280, 20226281, 20226282, 20226283, 20226284, 20226285, 20226286, 20226287, 20226288, 20226289, 20226290, 20226291, 20226292, 20240622, 20240623, 20240624, 20240625, 20240626, 20240627, 20240628, 20240629, 20240630, 20240631, 20240632, 20240633, 20240634, 20240635, 20240636, 20240637, 20240638, 20240639, 20240640, 20240641, 20240642, 20240643, 20240644, 20240645, 20240646, 20240647, 20240648, 20240649, 20240650, 20240651, 20240652, 20240653, 20240654, 20240655, 20240656, 20240657, 20240658, 20240659, 20240660, 20240661, 20240662, 20240663, 20240664, 20240665, 20240666, 20240667, 20240668, 20240669, 20240670, 20240671, 20240672, 20240673, 20240674, 20240675, 20240676, 20240677, 20240678, 20240679, 20240680, 20240681, 20240682, 20240683, 20240684, 20240685, 20240686, 20240687, 20240688, 20240689, 20240690, 20240691, 20254123, 20254124, 20254125, 20254126, 20254127, 20254128, 20254129, 20254130, 20254131, 20254132, 20254133, 20254134, 20254135, 20254136, 20254137, 20254138, 20254139, 20254140, 20254141, 20254142, 20254143, 20254144, 20254145, 20254146, 20254147, 20254148, 20254149, 20254150, 20254151, 20254152, 20254153, 20254154, 20254155, 20254156, 20254157, 20254158, 20254159, 20254160, 20254161, 20254162,…
    FDA recall number
    Z-1570-2025
    Quantity
    1661

    The text above is the beginning of the record; 15,209 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution in the states of WA, NE, VA, PA, DC, KY, CA, IL, WI, TX, MA, AR, MS, MI, OH, FL, NC, TN, LA, NJ, UT, NY, DE, CT, GA, IN, AZ, OK, MO, OR, MN, SC, NM, MD, IA, KS, AL, CO, ME and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CHILE, CZECH REPUBLIC, FINLAND, FRANCE, GEORGIA, GERMANY, HONG KONG, ISRAEL, ITALY, JAPAN, KUWAIT, LEBANON, LITHUANIA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, POLAND, PORTUGAL, ROMANIA, SERBIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UK, UNITED ARAB EMIRATES.

Source

Issued by
U.S. Food and Drug Administration
Firm
Thoratec LLC
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Thoratec LLC, Month: April 2025, Week 2025-W17. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Left Ventricular Assist System Mobile Power Unit (MPU) may have electrical component that may cause MPU not to turn on, unprompted shut down, or turning off/restarting, may cause system controller to indicate Yellow Wrench or No External Power alarms. If MPU losses power, System Controller can support the pump up to 15 minutes, and if not switched to rechargeable batteries, pump may lose power.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 96471 on April 23, 2025.