What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
16 more codes are in the record below.
HeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
- Lot or code
- PAD-PAK-01, Lot: H017-002-174 and H047-002-100, UDI: 0506016712046; PAD-PAK-02, Lot: H023-002-051, UDI: 0506016712047; PAD-PAK-03, Lot: H017-002-173, H047-002-101, UDI: 0506016712123; PAD-PAK-04, Lot: H023-002-050, UDI: 0506016712124; PAD-PAK-07, Lot: H017-002-178, UDI: 0506016712448; PAD-PAK-03J, Lot: H017-002-071 and H047-002-035, UDI: 0506016712475; PAD-PAK-04J, Lot: H023-002-009, UDI: 0506016712476
- FDA recall number
- Z-0288-2026
- Quantity
- 1,291,165 units
Distribution (FDA): US Nationwide distribution. Worldwide distribution to Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Latvia, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- HeartSine Technologies Ltd
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.