HeartSine Technologies Ltd recalls
4 notices naming HeartSine Technologies Ltd as the recalling firm, since August 6, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
HeartSine Technologies Ltd: HeartSine samaritan PAD
Hazard (FDA): Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
FDAMedical devices
HeartSine Technologies Ltd: HeartSine SAM 350P 350P US Device
Hazard (FDA): Due to a damaged (over stretched) membrane tail that causes an intermittent fault.
FDAMedical devices
HeartSine Technologies Ltd: HeartSine Pad-Pak
Hazard (FDA): Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
FDAMedical devices
HeartSine Technologies Ltd: HeartSine SAM 350P
Hazard (FDA): Due to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
FDAMedical devices