What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Optiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, (4) 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM
- Lot or code
- 1. Model: 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, UDI-DI: 840861100149, All Lots manufactured after SEP/06/2023, 2. Model: 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, UDI-DI: 840861100156, All Lots manufactured after SEP/06/2023, 3. Model: 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, UDI-DI: 840861100163, All Lots manufactured after SEP/06/2023, 4. Model: 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM, UDI-DI: 840861100170, All Lots manufactured after SEP/06/2023;
- FDA recall number
- Z-0041-2026
- Quantity
- 172,494,732 units (14,374,561 cases @ 12 units per case)
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Fresenius Medical Care Holdings, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.