What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Bicarby Dialysate; Model number: RFP-400-G;
- Lot or code
- Model number: RFP-400-G; UDI-DI: 00840861102839; Lot codes: 25SG06026, 25SG06027;
- FDA recall number
- Z-2401-2026
- Quantity
- 4752
Distribution (FDA): Domestic: TX.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Fresenius Medical Care Holdings, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.