What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300
- Lot or code
- List Number: 03P88-25. All lot numbers.
- FDA recall number
- Z-2583-2025
- Quantity
- 8,962,450 units
i-STAT EG7+ cartridge. List Number: 03P76-25.
- Lot or code
- List Number: 03P76-25. All lot numbers.
- FDA recall number
- Z-2584-2025
- Quantity
- 6,115,950 units
i-STAT EG6+ cartridge. List Number: 03P77-25.
- Lot or code
- List Number: 03P77-25. All lot numbers.
- FDA recall number
- Z-2585-2025
- Quantity
- 547,250 units
Distribution (FDA): US Nationwide distribution.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Abbott Point Of Care Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.