Abbott Point Of Care Inc. recalls
3 notices naming Abbott Point Of Care Inc. as the recalling firm, since September 24, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Abbott Point Of Care Inc.: i-STAT EG7+ cartridge and 2 more
Hazard (FDA): Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than…
FDAMedical devices
Abbott Point Of Care Inc.: i-STAT CG8+ cartridge. List Number: 03P88-25. 510 and 2 more
Hazard (FDA): Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
FDAMedical devices
Abbott Point Of Care Inc.: i-STAT CG4+ cartridge
Hazard (FDA): Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.
FDAMedical devices