FDA recall 97217: Philips Ultrasound, Inc: 3D6-2 Transducer Probe and 68 more

FDA recall 97217 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
Text from FDA, unedited.

FDA classification: Class III. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

391 more codes are in the record below.

  • 3D6-2 Transducer Probe

    Lot or code
    Model No. 989605440872, 8500-2032-01; UDI: N/A; Serial No. 031XK6, 02VF0G, 02M3N3, 031VQT, 03GKDF, 02TWD3, 02NDCL, 02CBD4, 02KWQ9.
    FDA recall number
    Z-2346-2025
    Quantity
    171,322 units
  • 3D9-3v Transducer Probe

    Lot or code
    Model No. 989605387181, 989605389272, 989605395621, 989605395622, 989605409562, 989605412042, 989605412082, 989605415031, 989605415032, 989605418972, 989605437982, 989605438451, 989605440832, 989606714071, 989606717071, 989606750021, 8500-1715-01; UDI: Unique Device Identifier (UDI), (01)00884838061545(21)B3HM2J, (01)00884838061545(21)B3FH50, 010088483806154521F0G00P, 010088483806154521F0G1J2, (01)00884838061545(21)B2LN6D, (01)00884838061545(21)B2NP6J, (01)00884838061545(21)F03YG9, (01)00884838061545(21)F03YMT, (01)00884838061545(21)F0JNY1, (01)00884838061545(21)F0JR9Z, (01)00884838061545(21)B2ZVMM, (01)00884838061545(21)F05MPP, (01)00884838061545(21)F0D8T6, (01)00884838061545(21)F0KF3X, (01)00884838061545(21)F0L24T, (01)00884838061545(21)F0L31F, (01)00884838061545(21)F0L31J, (01)00884838061545(21)F0L31N, 010088483806154521F0GBLQ, 010088483806154521F0GBLW, (01)00884838061545(21)F0K8CZ, (01)00884838061545(21)F0HM7P, (01)00884838061545(21)B286JT, (01)00884838061545(21)F0LLBD, (01)00884838061545(21)B24YHV, (01)00884838067707(21)B211VH, (01)00884838061545(21)F056HZ, (01)00884838061545(21)B39HPV, (01)00884838061545(21)B22GHQ, (01)00884838061545(21)F02H8B, 010088483806154521F02887, (01)00884838067707(21)B0NL3B, (01)00884838067707(21)B1TJ0C, (01)00884838061545(21)F0JNY0, (01)00884838061545(21)F0K02Z, (01)00884838061545(21)B35D7M, (01)00884838061545(21)F04286, (01)00884838061545(21)F056HR, (01)00884838061545(21)B2LMWK, (01)00884838061545(21)F04W16, (01)00884838061545(21)F05QVQ,…
    FDA recall number
    Z-2347-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 109,687 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • 7.5MHZ Endo Transducer Probe

    Lot or code
    UDI: N/A; Serial No. F00262.
    FDA recall number
    Z-2348-2025
    Quantity
    171,322 units
  • BP10-5ec Transducer Probe

    Lot or code
    Model No. 989605395642; UDI: (01)00884838061552(21)B29VPV, (01)00884838061552(21)B29VQ6, (01)00884838061552(21)B2CYJ4, (01)00884838061552(21)F06M63, (01)00884838061552(21)F06QYR, (01)00884838061552(21)B2CY7P, (01)00884838061552(21)F058R4, (01)00884838061552(21)B2CY8T, (01)00884838061552(21)F05YF6, (01)00884838061552(21)F0628C, (01)00884838061552(21)B29VPW, (01)00884838061552(21)F064N9, (01)00884838061552(21)F078HF, (01)00884838061552(21)F05Y6Y, (01)00884838061552(21)B2CYJJ, (01)00884838061552(21)B2CYD7; Serial No. B29VPV, B29VQ6, B2CYJ4, F06M63, F06QYR, B2CY7P, F058R4, B2CY8T, F05YF6, F0628C, B29VPW, F064N9, F078HF, F05Y6Y, B2CYJJ, B2CYD7.
    FDA recall number
    Z-2349-2025
    Quantity
    171,322 units
  • C10-3v Transducer Probe

    Lot or code
    Model No. 989605409562, 989605438521, 989605381661, 989605381663, 989605398231, 989605381662, 989605409561, 989605455171, 989605492481, 989605414531, 989605406772, 989605439232, 21781A, 989606708921, 989605406771, 989605414122; UDI: (01)00884838061583(21)B1ZPMG, (01)00884838061583(21)B3GL7D, (01)00884838061583(21)B3PJJ1, 010088483806158321F021FV, (01)00884838061583(21)B1R004, (01)00884838061583(21)B389HT, (01)00884838061583(21)B39YJ9, 010088483806158321F0GK85, 010088483806158321F0GK89, (01)00884838061583(21)B3C63X, (01)00884838061583(21)B3JDTZ, (01)00884838061583(21)B1QZX2, (01)00884838061583(21)F0LCKP, (01)00884838061583(21)B2K0BP, (01)00884838061583(21)B2NJN9, (01)00884838061583(21)F03XLP, (01)00884838061583(21)F0433X, (01)00884838061583(21)F055TN, (01)00884838061583(21)F05P66, (01)00884838061583(21)F0CK3H, (01)00884838061583(21)F0LFY4, (01)00884838061583(21)F0BJLL, (01)00884838061583(21)F0JL72, (01)00884838061583(21)F0JPYV, (01)00884838061583(21)B34LHZ, (01)00884838061583(21)F04VQ5, (01)00884838061583(21)F05HG5, (01)00884838061583(21)F05L7G, (01)00884838061583(21)F0D55B, (01)00884838061583(21)F0GQHD, (01)00884838061583(21)F0GVG0, (01)00884838061583(21)F0KRXK, (01)00884838061583(21)F088H3, (01)00884838061583(21)F08B16, (01)00884838061583(21)F08B18, (01)00884838061583(21)F09CL3, (01)00884838061583(21)F0JCQ0, (01)00884838061583(21)F031TM, (01)00884838067509(21)B1ZNPH, (01)00884838067509(21)B1ZP72, (01)00884838061583(21)B1QYCZ, (01)00884838061583(21)B39XR1,…
    FDA recall number
    Z-2350-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 254,348 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C10-4ec Transducer Probe

    Lot or code
    Model No. 989605413712, 989605438601, 989605391092; UDI: (01)00884838061507(21)B2KPHY, (01)00884838067837(21)B1JT51, (01)00884838067837(21)B1JTFN, (01)00884838067837(21)B1JTJY, (01)00884838067837(21)B1MT19, (01)00884838067837(21)B1MT9L, (01)00884838067837(21)B1MT9X, (01)00884838067837(21)B1MT9Y, (01)00884838067837(21)B1MTDR, (01)00884838067837(21)B1MTDX, (01)00884838067837(21)B1MTDZ, (01)00884838067837(21)B1MTY9, (01)00884838067837(21)B1N32V, (01)00884838067837(21)B1N33Y, (01)00884838067837(21)B1N340, (01)00884838067837(21)B1N34V, (01)00884838067837(21)B1N352, (01)00884838067837(21)B1N383, (01)00884838067837(21)B1N3CX, (01)00884838067837(21)B1N3FT, (01)00884838067837(21)B1N3MR, (01)00884838067837(21)B1N3MT, (01)00884838067837(21)B1N3V5, (01)00884838067837(21)B1RGK2, (01)00884838067837(21)B1RGL6, (01)00884838067837(21)B1RGL9, (01)00884838067837(21)B1RGR1, (01)00884838067837(21)B1RNJZ, (01)00884838067837(21)B1RNK5, (01)00884838067837(21)B1RNND, (01)00884838067837(21)B1RNQQ, (01)00884838067837(21)B1RNQR, (01)00884838067837(21)B1RP18, (01)00884838067837(21)B1RP19, (01)00884838067837(21)B1RP1C, (01)00884838067837(21)B1RP70, (01)00884838067837(21)B1WQZ1, (01)00884838067837(21)B1WRGN, (01)00884838067837(21)B1WRJC, (01)00884838067837(21)B1WYX1, (01)00884838067837(21)B1WYX3, (01)00884838067837(21)B1WYZ7, (01)00884838067837(21)B1WYZB, (01)00884838067837(21)B1WZ00, (01)00884838067837(21)B1WZ02, (01)00884838067837(21)B1WZ23, (01)00884838067837(21)B1WZ26, (01)00884838067837(21)B1WZ2G,…
    FDA recall number
    Z-2351-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 21,316 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • c5-1 Transducer Probe

    Lot or code
    Model No. 989605412042, 989605438541, 989605388071, 989605365003, 989605365001, 989605366331, 989605387852, 989605365004, 989605492481, 989605431531, 989605412072, 989606702251, 989605391092, 989605353303, 989605418552, 989605455171, 989605412041; UDI: (01)00884838061491(21)B2JTQ2, (01)00884838061507(21)B2KPHY, (01)00884838061613(21)B0TN11, (01)00884838061613(21)B0WPLV, (01)00884838061613(21)B0ZVFL, (01)00884838061613(21)B126WZ, (01)00884838061613(21)B13K7V, (01)00884838061613(21)B1434L, (01)00884838061613(21)B14QMX, (01)00884838061613(21)B16ZF8, (01)00884838061613(21)B17161, (01)00884838061613(21)B173X2, (01)00884838061613(21)B1743J, (01)00884838061613(21)B1811G, (01)00884838061613(21)B1C20X, (01)00884838061613(21)B1C5K5, (01)00884838061613(21)B1C764, (01)00884838061613(21)B1DT4H, (01)00884838061613(21)B1DXHY, (01)00884838061613(21)B1FDZ9, (01)00884838061613(21)B1K2G0, (01)00884838061613(21)B1K7F0, (01)00884838061613(21)B1LCZW, (01)00884838061613(21)B1LHJ3, (01)00884838061613(21)B1LHW3, (01)00884838061613(21)B1LNLZ, (01)00884838061613(21)B1LY8D, (01)00884838061613(21)B1LZW9, (01)00884838061613(21)B1M2VG, (01)00884838061613(21)B1M9J6, (01)00884838061613(21)B1M9QW, (01)00884838061613(21)B1M9QY, (01)00884838061613(21)B1M9RX, (01)00884838061613(21)B1MBPX, (01)00884838061613(21)B1MBQ3, (01)00884838061613(21)B1MKHL, (01)00884838061613(21)B1MKQM, (01)00884838061613(21)B1MKTB, (01)00884838061613(21)B1MKZF, (01)00884838061613(21)B1ML0V, (01)00884838061613(21)B1MNCG,…
    FDA recall number
    Z-2352-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 657,910 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C5-2 Transducer Probe

    Lot or code
    Model No. 989605358813, 989605371711, 989605371713, 989605386651, 989605387521, 989605389241, 989605389242, 989605398211, 989605399561, 989605418972, 989605430242, 989605451181, 989605454101, 989803002675; UDI: Unique Device Identifier (UDI), (01)00884838067424(21)B200XV, (01)00884838067424(21)B2W963, (01)00884838067424(21)B2XQ5R, (01)00884838067424(21)B31MDB, (01)00884838067424(21)B2BBY9, (01)00884838067424(21)B2C5FJ, (01)00884838067424(21)B329WV, (01)00884838067424(21)B3CWGT, (01)00884838067424, (01)00884838067424(21)F05G0L, (01)00884838061385(21)B1PKRD, (01)00884838067424(21)B224J4, (01)00884838067424(21)B224J7, (01)00884838067424(21)B224JX, (01)00884838067424(21)B224JZ, (01)00884838067424(21)B224MQ, (01)00884838067424(21)B2M756, (01)00884838067424(21)B2M75Q, (01)00884838067424(21)B2M776, (01)00884838067424(21)B34T1T, (01)00884838067424(21)B35Y9P, (01)00884838067424(21)B36LTF, (01)00884838068018(21)B24HQ5, (01)00884838067424(21)F07YGL, (01)00884838067424(21)B2HNC4, (01)00884838067943(21)B35H7Y, (01)00884838061385(21)B18L9Q, (01)00884838067424(21)B2L1Y0, (01)00884838067424(21)B2JX1Q, (01)00884838067424(21)F04LFN, (01)00884838068018(21)B27Q4P, (01)00884838068018(21)B27RCK, (01)00884838067943(21)B2PW77, (01)00884838067424(21)F011BB, (01)00884838068018(21)B27G1F, (01)00884838068018(21)B27ND4, (01)00884838068018(21)B27T3X, (01)00884838061385(21)B1TPP1, (01)00884838067424(21)B34T0G, (01)00884838061385(21)B1NMBP, (01)00884838061804(21)B1NLZK, (01)00884838061804(21)B1F3K0,…
    FDA recall number
    Z-2353-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 35,899 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C5-2 Lumify Transducer Probe

    Lot or code
    Model No. 989605450353, 989605450352, 989605477131, 989605450351, 989605451181, 989605427371; UDI: 010088483808517621F0FYRG, (01)00884838085176(21)B3H57K, (01)00884838068766(21)B1W1CR, (01)00884838086555(21)B2C7Z5, (01)00884838085176(21)F07VCW, (01)00884838085176(21)F0J0P8, (01)00884838061828(21)B1MBZM, (01)00884838068766(21)B1W1NG, (01)00884838085176(21)B1MC3H, (01)00884838085176(21)B1QHPM, (01)00884838085176(21)B1W13D, (01)00884838085176(21)B1W153, (01)00884838085176(21)B27BC5, (01)00884838085176(21)B3PH0F, (01)00884838085176(21)B3R48C, (01)00884838085176(21)F02N98, (01)00884838085176(21)F07PWD, (01)00884838085176(21)F07TZ0, (01)00884838085176(21)F07Y64, (01)00884838085176(21)F081Y3, (01)00884838085176(21)F082NG, (01)00884838085176(21)F082NP, (01)00884838085176(21)F0855V, (01)00884838085176(21)F088RF, (01)00884838085176(21)F089MZ, (01)00884838085176(21)F089N6, (01)00884838085176(21)F08GBG, (01)00884838085176(21)F08GBX, (01)00884838085176(21)F08HWH, (01)00884838085176(21)F08KFK, (01)00884838085176(21)F08NKZ, (01)00884838085176(21)F08NL0, (01)00884838085176(21)F08NLH, (01)00884838085176(21)F08PR7, (01)00884838085176(21)F08PR8, (01)00884838085176(21)F08PR9, (01)00884838085176(21)F08PRB, (01)00884838085176(21)F08PRF, (01)00884838085176(21)F08WGJ, (01)00884838085176(21)F08ZVL, (01)00884838085176(21)F091X3, (01)00884838085176(21)F091X6, (01)00884838085176(21)F091X8, (01)00884838085176(21)F091X9, (01)00884838085176(21)F092MC, (01)00884838085176(21)F092MG,…
    FDA recall number
    Z-2354-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 91,514 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C6-2 Transducer Probe

    Lot or code
    Model No. 989605418982, 989605418983, 989605463391, 989605463391, 989605442151, 989605387071, 989605418972; UDI: (01)00884838067868(21)B1N309, (01)00884838067868(21)B1N3GM, (01)00884838067868(21)B20LH9, (01)00884838087163(21)B27M92, (01)00884838087163(21)B2D4X5, (01)00884838087163(21)B2TQDL, (01)00884838087163(21)B2TQHL, (01)00884838087163(21)B2TR9L, (01)00884838087163(21)B2TR9R, (01)00884838087163(21)B32NLQ, (01)00884838087163(21)B32P57, (01)00884838087163(21)B3WW8C, (01)00884838087163(21)B40ZK4, (01)00884838087163(21)F05J29, (01)00884838087163(21)F0GGQZ, (01)04719873000036(21)B2TR3Y, (01)04719873000036(21)B3HZ2G, (01)04719873000036(21)F08334, (01)04719873000036(21)F0H6X0, 010088483808716321B3XNQK, 010471987300003621B2D52G, (01)00884838067868(21)B27KQ5, (01)00884838087163(21)F08BN7, (01)00884838067868(21)B1N327, (01)00884838067868(21)B1N35V, (01)00884838087163(21)B2THXY, (01)00884838087163(21)B2TJ1M, (01)00884838087163(21)B2ZMG6, (01)04719873000036(21)B3TKBZ, (01)04719873000036(21)B3W5VV, (01)04719873000036(21)F080L2, (01)00884838067868(21)B1WQXH, (01)00884838067868(21)B1WQZ5, (01)00884838067868(21)B1WR33, (01)00884838087163(21)B32Q2J, (01)00884838067868(21)B27KCV, (01)00884838067868(21)B27KNX, (01)00884838067868(21)B1N300, (01)00884838087163(21)F057NM, (01)00884838087163(21)F0HX1L, (01)00884838087163(21)B2PRZ1, (01)00884838067868(21)B24LLK, (01)00884838087163(21)B32NK1, (01)00884838067868(21)B20CNY, (01)00884838087163(21)B3W6RH, (01)00884838087163(21)F0552K,…
    FDA recall number
    Z-2355-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 75,225 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C6-3 Transducer Probe

    Lot or code
    Model No. 989605359593, 989605387371, 989605359594, 989605359591; UDI: (01)00884838067950(21)B20N4N, (01)00884838067950(21)B1WR9H, (01)00884838067950(21)B1RGN0, (01)00884838067950(21)B1MTT4, (01)00884838067950(21)B1N3C4, (01)00884838087170(21)B27LPZ, (01)00884838087170(21)B3W5CM, (01)00884838067950(21)B24KL3, (01)00884838067950(21)B1MTT5, (01)00884838087170(21)B27LQ2; Serial No. B20N4N, B1WR9H, B1RGN0, B1MTT4, B16QBM, B1N3C4, B27LPZ, 03DG3C, B3W5CM, B2ZN6B, B19M3K, B2TJFT, B35M7F, B24KL3, B1MTT5, B32P04, B35MC6, B27LQ2.
    FDA recall number
    Z-2356-2025
    Quantity
    171,322 units
  • C8-4v Transducer Probe

    Lot or code
    Model No. 989803002685, 989605416131, 989605408153, 989605430232, 989605416141, 989605364943, 989605388111, 989605387261, 989605408151, 989605361973, 989605408141, 989605430242, 989605364941; UDI: (01)00884838067745(21)B1YRNV, (01)00884838067981(21)B35BZF, (01)00884838067981(21)B24PBY, (01)00884838067981(21)B27T4Z, (01)00884838067981(21)B35H6L, (01)00884838067981(21)B1RPRM, (01)00884838068032(21)B2DGF2, (01)00884838067981(21)B1MZ1D, (01)00884838067981(21)B1X19N, (01)00884838067981(21)B1MYCZ, (01)00884838067981(21)B1RKT8, (01)00884838067981(21)B3PYXD, (01)00884838067981(21)B2TL3C, (01)00884838067981(21)B20NKJ, (01)00884838068032(21)B1X2X3, (01)00884838068032(21)B2TN85, (01)00884838068032(21)B2TTB8, (01)00884838068032(21)B2ZHBM, (01)00884838068032(21)B2ZHDD, (01)00884838067981(21)B2TKV5, (01)00884838067981(21)B40GX1, (01)00884838068032(21)B1RRJL, (01)00884838067981(21)B3ZCR0, (01)00884838067981(21)F09JPD, (01)00884838067981(21)B20G80, (01)00884838067981(21)B2PKDF, (01)00884838067981(21)B2TL8Y, (01)00884838067981(21)B20NHL, (01)00884838064577(21)B1YRVW, (01)00884838067981(21)B1RQDW, (01)00884838067981(21)B24FWM, (01)00884838067981(21)B2PKT2, (01)00884838067981(21)B35P3D, (01)00884838067981(21)F08W6H, (01)00884838067981(21)B35C6P, (01)00884838067981(21)B1MXNG, (01)00884838067981(21)B20HV9, (01)00884838064577(21)B1F3XP, (01)00884838064577(21)B1F3RL, (01)00884838067981(21)B1MXCF, (01)00884838067981(21)B1RPML, (01)00884838067981(21)B39MMX, (01)00884838067981(21)B39RBC,…
    FDA recall number
    Z-2357-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 24,841 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C8-4v Transducer Probe

    Lot or code
    Model No. 989803002685, 989605416131, 989605408153, 989605430232, 989605416141, 989605364943, 989605388111, 989605387261, 989605408151, 989605361973, 989605408141, 989605430242, 989605364941; UDI: Unique Device Identifier (UDI), (01)00884838067745(21)B1YRNV, (01)00884838067981(21)B35BZF, (01)00884838067981(21)B24PBY, (01)00884838067981(21)B27T4Z, (01)00884838067981(21)B35H6L, (01)00884838067981(21)B1RPRM, (01)00884838068032(21)B2DGF2, (01)00884838067981(21)B1MZ1D, (01)00884838067981(21)B1X19N, (01)00884838067981(21)B1MYCZ, (01)00884838067981(21)B1RKT8, (01)00884838067981(21)B3PYXD, (01)00884838067981(21)B2TL3C, (01)00884838067981(21)B20NKJ, (01)00884838068032(21)B1X2X3, (01)00884838068032(21)B2TN85, (01)00884838068032(21)B2TTB8, (01)00884838068032(21)B2ZHBM, (01)00884838068032(21)B2ZHDD, (01)00884838067981(21)B2TKV5, (01)00884838067981(21)B40GX1, (01)00884838068032(21)B1RRJL, (01)00884838067981(21)B3ZCR0, (01)00884838067981(21)F09JPD, (01)00884838067981(21)B20G80, (01)00884838067981(21)B2PKDF, (01)00884838067981(21)B2TL8Y, (01)00884838067981(21)B20NHL, (01)00884838064577(21)B1YRVW, (01)00884838067981(21)B1RQDW, (01)00884838067981(21)B24FWM, (01)00884838067981(21)B2PKT2, (01)00884838067981(21)B35P3D, (01)00884838067981(21)F08W6H, (01)00884838067981(21)B35C6P, (01)00884838067981(21)B1MXNG, (01)00884838067981(21)B20HV9, (01)00884838064577(21)B1F3XP, (01)00884838064577(21)B1F3RL, (01)00884838067981(21)B1MXCF, (01)00884838067981(21)B1RPML, (01)00884838067981(21)B39MMX,…
    FDA recall number
    Z-2358-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 24,871 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C8-5 Transducer Probe

    Lot or code
    Model No. 989605391922, 989605438441, 989605440822, 989605387852, 989605387281, 989605391092, 8500-1664-01, 989605352343, 989605352342; UDI: (01)00884838061514(21)B1Q41K, (01)00884838061514(21)B1QB9Z, (01)00884838061514(21)B2859G, (01)00884838061514(21)B2YTR3, (01)00884838061514(21)B2YTTV, (01)00884838061514(21)B3F4JN, (01)00884838061514(21)F08DZC, 010088483806151421F01W6L, (01)00884838061514(21)F04XX6, (01)00884838061514(21)F0FCL7, (01)00884838061514(21)B3B4RD, (01)00884838061514(21)B288HM, (01)00884838061514(21)B2FXYD, (01)00884838061514(21)B2FXZB, (01)00884838061514(21)B1P4JM, (01)00884838061514(21)B1P9ZG, (01)00884838061514(21)B338L3, (01)00884838061514(21)F0C7VZ, (01)00884838061514(21)F0FHNV, (01)00884838061514(21)F0KM0Y, (01)00884838061514(21)B22GYC, (01)00884838061514(21)B21T06, (01)00884838067691(21)B26NBY, (01)00884838061514(21)B34J00, (01)00884838061514(21)B2FXJK, (01)00884838067691(21)B2882D, (01)00884838067691(21)B288VM, (01)00884838067691(21)B2FY2G, (01)00884838061514(21)B2YH6Z, (01)00884838061514(21)B30Q8X, 010088483806151421F0G4ZF, (01)00884838061514(21)F011BM, (01)00884838061514(21)B1QBW2, (01)00884838061514(21)B1QTN7, (01)00884838061514(21)B1QTVX, (01)00884838061514(21)B2YGHZ, (01)00884838061514(21)F018FW, (01)00884838061514(21)B37W7W, (01)00884838061514(21)B22NLJ, (01)00884838067691(21)B338F0, (01)00884838061514(21)B216LD, (01)00884838061514(21)F0B6W5, (01)00884838061514(21)B287ZV, (01)00884838061514(21)B288BD, (01)00884838061514(21)B29XXK,…
    FDA recall number
    Z-2359-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 48,158 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C9-2 Transducer Probe

    Lot or code
    Model No. 989605391092, 989605438431, 989605418552, 989605391091, 989605412042; UDI: (01)00884838061507(21)B3G8WQ, 010088483806150721F0G8PX, (01)00884838061507(21)B25KL1, (01)00884838061507(21)B2L2LB, 010088483806150721F0FWFW, 010088483806150721F0G0CF, (01)00884838061507(21)F02CP6, (01)00884838061507(21)B26BXV, (01)00884838061507(21)B28G5L, (01)00884838061507(21)B2FG9B, (01)00884838061507(21)B3G8M5, (01)00884838061507(21)F03V2C, (01)00884838061507(21)F040X0, (01)00884838061507(21)F0G485, (01)00884838061507(21)B3NLL9, (01)00884838061507(21)F03W2J, 010088483806150721F0G524, (01)00884838061507(21)B1ZXP6, (01)00884838061507(21)F03Y53, (01)00884838061507(21)F0860Z, (01)00884838061507(21)F0JPXD, (01)00884838061507(21)B331YR, (01)00884838061507(21)F05CDM, (01)00884838061507(21)F05DYV, (01)00884838061507(21)F05VLL, 010088483806150721F0G59Y, (01)00884838061507(21)B3RFXC, (01)00884838061507(21)F03TK9, (01)00884838061507(21)B22HP1, (01)00884838061507(21)B250V0, (01)00884838061507(21)B1PL8N, (01)00884838061507(21)B3MHJM, (01)00884838061507(21)B2FQDH, (01)00884838061507(21)F060VV, (01)00884838061507(21)F0H3CD, (01)00884838061507(21)B1Q584, (01)00884838061507(21)B1W051, 010088483806150721F0299F, (01)00884838061507(21)B1NW7V, (01)00884838061507(21)F0KL3Q, (01)00884838061507(21)B3JTJX, (01)00884838061507(21)F0LK5V, (01)00884838061507(21)F07XGB, (01)00884838061507(21)F080X4, (01)00884838061507(21)F0818R, (01)00884838061507(21)F09B0J, (01)00884838061507(21)B2FQPR, (01)00884838061507(21)B2KP1Y,…
    FDA recall number
    Z-2360-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 119,237 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C9-3io Transducer Probe

    Lot or code
    Model No. 989605395612; UDI: (01)00884838061538(21)B2CYBN, (01)00884838061538(21)B0WM4Z, (01)00884838061538(21)B0WM4P; Serial No. B2CYBN, B0WM4Z, B0WM4P.
    FDA recall number
    Z-2361-2025
    Quantity
    171,322 units
  • C9-3v Transducer Probe

    Lot or code
    Model No. 989605366341; UDI: (01)00884838024304(21)B0LBJX, (01)00884838024304(21)B0LBXV; Serial No. B0DCXF, B0CT9W, 038VLZ, B0LBY2, 037RDB, B0DD3B, B0LBJX, B0LBXV, 037RG6, B0JDHK, B0BMR8.
    FDA recall number
    Z-2362-2025
    Quantity
    171,322 units
  • C9-4 Transducer Probe

    Lot or code
    Model No. 989605365021, 989605365023, 989605387291; UDI: (01)00884838064607(21)B1FZ1B, (01)00884838064607(21)B2KL13, (01)00884838064607(21)B30GDB, (01)00884838064607(21)B33HMW, (01)00884838064607(21)B3CL5L, (01)00884838064607(21)B1VKTB, (01)00884838064607(21)B2KL31, (01)00884838064607(21)B2JZ44, (01)00884838064607(21)B3GX1T, (01)00884838064607(21)B1XRF3, (01)00884838064607(21)B34QB8, (01)00884838064607(21)B1GFMQ, (01)00884838064607(21)B1RYG9, (01)00884838064607(21)B23TC6, (01)00884838064607(21)B1XRF1, (01)00884838064607(21)B1XN5J, (01)00884838064607(21)B1RYG6, (01)00884838064607(21)B2JTTX, (01)00884838064607(21)B3DLZ0, (01)00884838064607(21)B33XZB, (01)00884838064607(21)B1NPG7, (01)00884838064607(21)B1VJYF, (01)00884838064607(21)B3H0QV, (01)00884838064607(21)B3LNZH, (01)00884838064607(21)B31DH7, (01)00884838064607(21)B1RVCQ, (01)00884838064607(21)B34VJ7, (01)00884838064607(21)B3DC5D, (01)00884838064607(21)B1RYL2, (01)00884838064607(21)B1RVCR, (01)00884838064607(21)B1RYN1, (01)00884838064607(21)B1VJYH, (01)00884838064607(21)B31DH4, (01)00884838064607(21)B1MQJ5, (01)00884838064607(21)B23T6T, (01)00884838064607(21)B1RYJP, (01)00884838064607(21)B1ZRQM, (01)00884838064607(21)B1RYMZ, (01)00884838064607(21)B1QBG1, (01)00884838064607(21)B33XZD, (01)00884838064607(21)B3BWZ4, (01)00884838064607(21)B2KL87, (01)00884838064607(21)B3DLZ3, (01)00884838064607(21)B1RT45; Serial No. B1FZ1B, B2KL13, B30GDB, B33HMW, B3CL5L, B1VKTB, B2KL31, B2JZ44, B3GX1T, B1XRF3, B34QB8, B1GFMQ, B1F22H, B1RYG9,…
    FDA recall number
    Z-2363-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 288 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C9-4v Transducer Probe

    Lot or code
    Model No. 989605389252, 989605418972, 989605463411, 989605442141, 989605463411, 989605389251, 989605412042, 989605387091, 989605418971; UDI: (01)00884838061392(21)B1Q4CD, (01)00884838061392(21)B1YXKM, (01)00884838061392(21)B323T7, (01)00884838061392(21)F00QWD, (01)00884838061392(21)F029RH, (01)00884838067875(21)B1N3PG, (01)00884838067875(21)B1N3PZ, (01)00884838067875(21)B1N3TV, (01)00884838067875(21)B1N3W3, (01)00884838067875(21)B1N3ZJ, (01)00884838067875(21)B20M3Y, (01)00884838067875(21)B27LZD, (01)00884838067875(21)B27LZH, (01)00884838067875(21)B2TR7C, (01)00884838067875(21)B2ZMR5, (01)00884838067875(21)B32NH8, (01)00884838067875(21)B39VT0, (01)00884838067875(21)B410QW, (01)00884838067875(21)F05HKB, (01)00884838067875(21)F0JQQG, (01)04719873000050(21)B2PTCD, (01)04719873000050(21)B32PHY, 010088483806787521B3XMR5, N/A, N/A, N/A, N/A, N/A, N/A, N/A, (01)00884838067875(21)F0L3Z8, (01)00884838061392(21)B30WYR, (01)00884838067875(21)B3Z7ZM, (01)00884838061392(21)B232FN, (01)00884838061392(21)B25HKF, (01)00884838061392(21)B3LLMF, (01)00884838061392(21)B1WF1V, (01)00884838061392(21)B25JBQ, (01)00884838061392(21)B26BTD, (01)00884838061392(21)F07X9C, (01)00884838061392(21)F089ZV, (01)00884838067875(21)B1WZDW, (01)00884838067875(21)F08RGM, (01)00884838061392(21)B1ZV0G, (01)00884838061392(21)B1ZV0K, (01)00884838061392(21)B1ZV0M, (01)00884838061392(21)B1ZV2W, (01)00884838061392(21)B20233, (01)00884838061392(21)B2LTD7, (01)00884838061392(21)B2LTMJ, (01)00884838061392(21)B2LTP6,…
    FDA recall number
    Z-2364-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 78,589 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • C9-5ec Transducer Probe

    Lot or code
    Model No. 989605387351, 989605440802; UDI: (01)00884838067677(21)B1X88J, (01)00884838067677(21)B264HW, (01)00884838067677(21)B36JG7, (01)00884838067677(21)B29HQQ, (01)00884838067677(21)B264LH, (01)00884838067677(21)B2X78L, (01)00884838067677(21)B3055N, (01)00884838067677(21)B284J2, (01)00884838067677(21)F06LP2, (01)00884838067677(21)B1K3BP, (01)00884838067677(21)B32262, (01)00884838067677(21)B36JHC, (01)00884838067677(21)B2J631, (01)00884838067677(21)B3224T, (01)00884838067677(21)F02K00, (01)00884838067677(21)B2X710, (01)00884838067677(21)B3MDX4, (01)00884838067677(21)B1LJYQ, (01)00884838067677(21)B36JPK, (01)00884838067677(21)F02R0G, (01)00884838067677(21)B3225C, (01)00884838067677(21)B1M76Q, (01)00884838067677(21)B1X891, (01)00884838067677(21)B1X8BM, (01)00884838067677(21)B1X8CY, (01)00884838067677(21)B1X8FQ, (01)00884838067677(21)B201N5, (01)00884838067677(21)B201N7, (01)00884838067677(21)B29JBL, (01)00884838067677(21)B2X71C, (01)00884838067677(21)B2X7BN, (01)00884838067677(21)B29HTL, (01)00884838067677(21)B2X71M, (01)00884838067677(21)F02LY6, (01)00884838067677(21)B29HX3, (01)00884838067677(21)F04F0Z, (01)00884838067677(21)F0B377, (01)00884838067677(21)F0BBHC, (01)00884838067677(21)B1FRLC, (01)00884838067677(21)F0568P, (01)00884838067677(21)B1M6XC, (01)00884838067677(21)B1M6X7, (01)00884838067677(21)B1M6Z0, (01)00884838067677(21)B1M6W4, (01)00884838067677(21)B264NN, (01)00884838067677(21)B32288, (01)00884838067677(21)B1M733, (01)00884838067677(21)B1X8C6,…
    FDA recall number
    Z-2365-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 2,005 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • D2cwc Transducer Probe

    Lot or code
    Model No. 989605438551, 989605412052, 989605400802; UDI: (01)00884838061347(21)B1XJV9, (01)00884838061347(21)B1XKWM, (01)00884838061347(21)B1YLM8, (01)00884838061347(21)B2RY5Q, (01)00884838061347(21)B33Z0G, (01)00884838061347(21)B33Z8X, (01)00884838061347(21)F07Y3Y, (01)00884838061347(21)F0DB0C, 010088483806134721F0GCXL, 010088483806134721F0G55C, N/A, (01)00884838061347(21)B1YLVZ, (01)00884838061347(21)B2YJM2, (01)00884838061347(21)F09VWC, N/A, (01)00884838061347(21)F052VD, (01)00884838061347(21)F087H0, (01)00884838061347(21)F08LH8, (01)00884838061347(21)F08WX7, (01)00884838061347(21)F0FBP8, N/A, (01)00884838061347(21)B1XKC3, (01)00884838061347(21)B1YKQL, (01)00884838061347(21)B1YLM2, (01)00884838061347(21)B1YLQJ, (01)00884838061347(21)B1YKJC, (01)00884838061347(21)B1XH5T, (01)00884838061347(21)B33YV3, (01)00884838061347(21)F031Z6, (01)00884838061347(21)F0D039, (01)00884838061347(21)B1XGDD, (01)00884838061347(21)B2YJ5F, (01)00884838061347(21)B2YJD7, (01)00884838061347(21)B33ZW9, (01)00884838061347(21)B33ZWX, (01)00884838061347(21)B33ZY3, (01)00884838061347(21)F02CMG, (01)00884838061347(21)F02JZ3, (01)00884838061347(21)F0F8T7, (01)00884838061347(21)F0FD89, N/A, N/A, (01)00884838061347(21)F042XV, (01)00884838061347(21)B2RXX3, 010088483806134721F01RMH, (01)00884838061347(21)F01Q38, (01)00884838061347(21)F0L5LJ, (01)00884838061347(21)F0L5LK, (01)00884838061347(21)F0LLW2, (01)00884838061347(21)B1YLZP, (01)00884838061347(21)F0CH17, (01)00884838061347(21)F0D7Q9, N/A, N/A, N/A, N/A,…
    FDA recall number
    Z-2366-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 595,644 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • D2TCD Transducer Probe

    Lot or code
    Model No. 989605412102, 989605438591; UDI: (01)00884838061361(21)B1N4HT, (01)00884838061361(21)B1N4J6, (01)00884838061361(21)B2L6QZ, (01)00884838061361(21)B2L6R0, (01)00884838061361(21)B2L6T9, (01)00884838061361(21)F01C9Y, (01)00884838061361(21)F074NX, (01)00884838061361(21)B2Q7FK, (01)00884838061361(21)F08MVZ, (01)00884838061361(21)B1NDJD, (01)00884838061361(21)B1N4JF, (01)00884838061361(21)B2L6TV, (01)00884838061361(21)F05Z77, (01)00884838061361(21)B2Q71X, (01)00884838061361(21)F0998F, (01)00884838061361(21)F0998J, (01)00884838061361(21)F09WC4, (01)00884838061361(21)F02H8Q, (01)00884838061361(21)F07JPJ, (01)00884838061361(21)F00LFM, (01)00884838061361(21)F00LFN, (01)00884838061361(21)B2Q7FT, (01)00884838061361(21)B2L6RJ, (01)00884838061361(21)B2L6RF, (01)00884838061361(21)B1N4KJ, (01)00884838061361(21)F067C1, (01)00884838061361(21)F068GD, (01)00884838061361(21)B1N4JR, (01)00884838061361(21)B2Q7FZ, (01)00884838061361(21)F01J0W, (01)00884838061361(21)B2Q71V, (01)00884838061361(21)B2Q71W, (01)00884838061361(21)B1N4HG, (01)00884838061361(21)F01J0R, (01)00884838061361(21)F0BBGN, (01)00884838061361(21)F0BFWZ, (01)00884838061361(21)F00436, (01)00884838061361(21)F00K75, (01)00884838061361(21)F00K76, (01)00884838061361(21)B2L6QD, (01)00884838061361(21)B2L6QW, (01)00884838061361(21)B2L6R1, (01)00884838061361(21)F04KMV, (01)00884838061361(21)F04KMW, (01)00884838061361(21)F04LKF, (01)00884838061361(21)B2L6TW, (01)00884838061361(21)B1N4JK, (01)00884838061361(21)B2L6T1,…
    FDA recall number
    Z-2367-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 3,171 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • D5cwc Transducer Probe

    Lot or code
    Model No. 989605438561, 989605412062; UDI: (01)00884838061354(21)B1XLGQ, (01)00884838061354(21)B2L6LQ, (01)00884838061354(21)B1YGH9, (01)00884838061354(21)F0F143, (01)00884838061354(21)B1YGN1, (01)00884838061354(21)B2Q6W8, (01)00884838061354(21)B2L6MM, (01)00884838061354(21)B2Q6PV, (01)00884838061354(21)B2Q6QF, (01)00884838061354(21)F0FYTZ, (01)00884838061354(21)F04N02, (01)00884838061354(21)B1YGJH, (01)00884838061354(21)B1YGJK, (01)00884838061354(21)B2L6LH, (01)00884838061354(21)B2L6LT, (01)00884838061354(21)F0D4X3, (01)00884838061354(21)B1N4NC, (01)00884838061354(21)B2Q6CZ, (01)00884838061354(21)B1YGHW, (01)00884838061354(21)B2Q6MJ, (01)00884838061354(21)B2Q6ML, (01)00884838061354(21)F0367V, (01)00884838061354(21)F0367Y, (01)00884838061354(21)B1XL7W, (01)00884838061354(21)B1XLF6, (01)00884838061354(21)B2L6LZ, (01)00884838061354(21)B2Q6JJ, (01)00884838061354(21)F07TXW, (01)00884838061354(21)F05L2W, (01)00884838061354(21)B1XLDJ, (01)00884838061354(21)F07XDK, (01)00884838061354(21)B1YGM2, (01)00884838061354(21)F08MVP, (01)00884838061354(21)F0F839, (01)00884838061354(21)F0F83D, (01)00884838061354(21)F0F8TP, (01)00884838061354(21)F02WL4, (01)00884838061354(21)B1XLH1, (01)00884838061354(21)B1YGMH, (01)00884838061354(21)B1YGMV, (01)00884838061354(21)B2Q6DF, (01)00884838061354(21)B1XLFC, (01)00884838061354(21)B1XLBL, (01)00884838061354(21)B1YGNB, (01)00884838061354(21)B2Q6HH, (01)00884838061354(21)B2Q6HK, (01)00884838061354(21)B2Q6HV, (01)00884838061354(21)F0F8TK,…
    FDA recall number
    Z-2368-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 12,053 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • eL18-4 Transducer Probe

    Lot or code
    Model No. 989605431531, 989605456291, 989605391092, 989605412062, 989605409562; UDI: (01)00884838070806(21)B3DVRN, (01)00884838070806(21)B3NTCY, (01)00884838070806(21)F05K3C, (01)00884838070813(21)B3J1RW, (01)00884838070813(21)B3KZY9, (01)00884838070813(21)F04T5F, (01)00884838070813(21)F0952L, (01)00884838070813(21)F0DYQG, (01)00884838070813(21)F0KLBR, 010088483807080621F01T01, (01)00884838070813(21)F0LWXK, (01)00884838070813(21)F0GMZ9, (01)00884838070806(21)F0JYQZ, (01)00884838070813(21)F0GN24, 010088483807081321F0FT37, 010088483807081321F0GD1D, 010088483807081321F0GDDJ, (01)00884838070806(21)B39ZR0, (01)00884838070806(21)B3BYRP, (01)00884838070806(21)B3KRZF, (01)00884838070813(21)F0LV0Z, (01)00884838070806(21)B3KRZG, (01)00884838070806(21)F04836, (01)00884838070806(21)F04N2M, (01)00884838070813(21)B3JQW5, (01)00884838070813(21)B3VGVT, (01)00884838070806(21)F03QB1, (01)00884838070806(21)F03XVR, (01)00884838070813(21)F089C4, (01)00884838070806(21)F0GTGK, (01)00884838070806(21)F0LDQ9, (01)00884838070806(21)F0BPTR, (01)00884838070806(21)F0JYQT, (01)00884838070806(21)F0JYR1, (01)00884838070806(21)B35Y87, (01)00884838070806(21)F0CXZH, (01)00884838070806(21)F0KW6H, (01)00884838070813(21)F05WMG, (01)00884838070813(21)F0F0KX, (01)00884838070813(21)F0GV7P, (01)00884838070806(21)F047WP, (01)00884838070806(21)F04D44, (01)00884838070806(21)F08XGX, (01)00884838070806(21)F0L6W6, (01)00884838070806(21)F031MC, (01)00884838070813(21)F0KPXQ, (01)00884838070813(21)F0KQZY,…
    FDA recall number
    Z-2369-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 278,593 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L10-4LAP Transducer Probe

    Lot or code
    Model No. 989605395602; UDI: (01)00884838061521(21)B0WM1V, (01)00884838061521(21)B0WM2G; Serial No. B0WM1V, B0WM2G.
    FDA recall number
    Z-2370-2025
    Quantity
    171,322 units
  • L11-3 Transducer Probe

    Lot or code
    Model No. 989605388251, 989605400511, 989605400513, 989605440762; UDI: (01)00884838067639(21)B1Y3CY, (01)00884838067967(21)F08RH4, (01)00884838067639(21)B1Y33W, (01)00884838067967(21)B1X2F6, (01)00884838067639(21)B1RY1G, (01)00884838067639(21)B1RXQ9, (01)00884838067639(21)B305V2, (01)00884838067967(21)B2H0PB, (01)00884838067967(21)B2PQ4H, (01)00884838067967(21)B2GZFN, (01)00884838067967(21)B2PV1V, (01)00884838067967(21)B2PTWK, (01)00884838067967(21)B2PMRG, (01)00884838067967(21)B2PMR8, (01)00884838067967(21)F08MFD, (01)00884838067967(21)B20DKD, (01)00884838067967(21)B24Q2T, (01)00884838067967(21)B27J4W, (01)00884838067967(21)B32RQX, (01)00884838067639(21)B1Y2DB, (01)00884838067967(21)B20KMM, (01)00884838067967(21)B1WTXZ, (01)00884838067967(21)B3YNDY, (01)00884838067639(21)B1Y3FG, (01)00884838067967(21)B1RMX9, (01)00884838067639(21)B2NN5M, (01)00884838067967(21)B2ZV3F, (01)00884838067639(21)B1Y3DR, (01)00884838067967(21)B24Q31, (01)00884838067967(21)F0GJN2, (01)00884838067967(21)B2PPZ7; Serial No. B1Y3CY, B14GMY, F08RH4, B1Y33W, B1X2F6, B1RY1G, B1RXQ9, B305V2, B2H0PB, B2PQ4H, B2GZFN, B2PV1V, B2PTWK, B2PMRG, B2PMR8, F08MFD, B19HW9, B20DKD, B24Q2T, B27J4W, B1CDP5, B32RQX, B19R86, B1Y2DB, B20KMM, B19H1P, B1WTXZ, B3YNDY, B1Y3FG, B0B20T, B1RMX9, B2NN5M, B1DCBR, B2ZV3F, B1Y3DR, B24Q31, F0GJN2, B2PPZ7.
    FDA recall number
    Z-2371-2025
    Quantity
    171,322 units
  • L12-3 Transducer Probe

    Lot or code
    Model No. 989605412072, 989605492481, 989605438571, 989605430253, 989605348243, 989605412071, 989605366211, 989605412042, 989605387241, 989605400521, 21475A, 989605456291; UDI: (01)00884838061620(21)B1WH6H, (01)00884838061620(21)B1WK9N, (01)00884838061620(21)B2278N, (01)00884838061620(21)B2ZC90, (01)00884838061620(21)B3GLHC, (01)00884838061620(21)B3KQWJ, (01)00884838061620(21)B3RB89, (01)00884838061620(21)F0112F, (01)00884838090804(21)B3F4N0, (01)00884838090804(21)B3NJZK, (01)00884838090804(21)F04ZGP, (01)00884838090804(21)F06W70, (01)00884838090804(21)F0B5FT, (01)00884838090804(21)F0D97N, (01)00884838090804(21)F0KNM8, 010088483809080421F021KZ, 010088483809080421F023DB, (01)00884838090804(21)F0LJNK, N/A, (01)00884838061620(21)B3BTG6, (01)00884838090804(21)F0K3YK, (01)00884838061620(21)F0KJT5, (01)00884838061620(21)B1QMN8, (01)00884838061620(21)B24X8V, (01)00884838061620(21)B24XKZ, 010088483809080421F0G2BJ, (01)00884838061620(21)B20XVR, (01)00884838061620(21)F06NV0, (01)00884838061620(21)F0CB3Z, (01)00884838090804(21)F0CX7C, (01)00884838061620(21)F0LB13, (01)00884838090804(21)B351B4, (01)00884838090804(21)B3JV93, (01)00884838061620(21)F04T3X, (01)00884838061620(21)F05CFK, (01)00884838090804(21)F012JG, (01)00884838061620(21)B2NX9G, (01)00884838061620(21)B2NXH7, (01)00884838061620(21)B2Q9B5, (01)00884838061620(21)B2YRFC, (01)00884838090804(21)B3JGWH, (01)00884838090804(21)B3JV8R, (01)00884838090804(21)F04HD2, (01)00884838090804(21)F04M2G, 010088483806162021F0293W,…
    FDA recall number
    Z-2372-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 585,791 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L12-4 Transducer Probe

    Lot or code
    Model No. 989605430652, 989605409562, 989605427122, 989605463401, 989605430651, 989605442161, 989605418963, 989605387852, 989605418701, 989605387081, 989605412042, 989605451171, 989605463401; UDI: (01)00884838061484(21)B1X9ZR, (01)00884838061484(21)B2Y9KT, (01)00884838061484(21)B30VBJ, (01)00884838061484(21)F001VD, (01)00884838061583(21)B1XBZR, (01)00884838061644(21)B1NM93, (01)00884838061644(21)B1NM97, (01)00884838061644(21)B1PKN3, (01)00884838061644(21)B1PKNX, (01)00884838061644(21)B1QC9Q, (01)00884838061644(21)B2C3WV, (01)00884838061644(21)B36X3B, (01)00884838061644(21)B3G91R, (01)04719873000043(21)B3Q2BY, (01)04719873000043(21)F0FFRZ, (01)04719873000043(21)F0HHBM, (01)04719873000043(21)F0JKV3, 010471987300004321B24KV9, (01)00884838067882(21)B32NGP, (01)00884838067882(21)B3FP9D, (01)00884838067882(21)B27L19, (01)00884838067882(21)B2D56Y, (01)00884838061491(21)B2JT8X, (01)00884838067882(21)B1N3XG, (01)00884838067882(21)B1N3Z3, (01)00884838067882(21)B2PT8J, (01)00884838067882(21)B2THVB, (01)00884838067882(21)B2THVK, (01)00884838067882(21)B2ZN60, (01)00884838067882(21)F0DYMD, (01)00884838067882(21)F0DYMR, 010088483806788221F0G3JP, 010088483806788221F0G869, 010088483806788221F0G86F, 010088483806788221F0GCX4, (01)00884838061644(21)B25L78, (01)00884838061644(21)B25L7X, (01)00884838067882(21)B3XP02, (01)00884838061644(21)B1WDL1, (01)00884838067882(21)F083L7, (01)00884838067882(21)F0JDWK, (01)00884838061484(21)B28GVF, (01)00884838061484(21)F071WF, (01)04719873000043(21)B3L8J2,…
    FDA recall number
    Z-2373-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 69,792 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L12-4 Lumify Transducer Probe

    Lot or code
    Model No. 989605434813, 989605451171, 989605434812, 989605477121, 989605434811; UDI: (01)00884838085169(21)B3GM1J, (01)00884838072305(21)B21GN4, (01)00884838085169(21)F07MZ5, (01)00884838068759(21)B21Z05, (01)00884838085169(21)B35YH3, (01)00884838085169(21)F0FG0Y, (01)00884838086586(21)B2C7QM, (01)00884838085169(21)B25WBT, (01)00884838085169(21)F01RBW, (01)00884838085169(21)F0CJ0M, (01)00884838085169(21)F0KW5F, (01)00884838085169(21)F00CDB, (01)00884838085169(21)F0DC9D, (01)00884838085169(21)F0DCLR, (01)00884838085169(21)F0DDF7, (01)00884838085169(21)F0DF57, (01)00884838085169(21)F0DF5D, (01)00884838085169(21)F02WK6, (01)00884838085169(21)F08PRN, (01)00884838085169(21)B2J24P, (01)00884838086586(21)B2C7PX, 010088483808516921B27DHQ, 010088483808516921B2J2CN, (01)00884838085169(21)F0F0B8, (01)00884838085169(21)F0FCP7, (01)00884838085169(21)B27CFY, (01)00884838085169(21)B27F3B, (01)00884838085169(21)F066CP, (01)00884838085169(21)F00PYH, (01)00884838085169(21)F0KTDZ, (01)00884838085169(21)F0J2M9, (01)00884838085169(21)B1L75L, (01)00884838085169(21)F037QC, (01)00884838085169(21)F07R0D, (01)00884838085169(21)F03FL7, (01)00884838085169(21)F0BMD1, (01)00884838061835(21)B1NJMW, (01)00884838061835(21)B1NJP2, (01)00884838061835(21)B1NJQZ, (01)00884838061835(21)B1NJR9, (01)00884838068759(21)B1QGV5, (01)00884838068759(21)B1QH43, (01)00884838068759(21)B1QH4K, (01)00884838068759(21)B1QH4Y, (01)00884838068759(21)B1QH50, (01)00884838068759(21)B1QH61, (01)00884838068759(21)B1QH6F,…
    FDA recall number
    Z-2374-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 270,814 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L12-5 38mm Transducer Probe

    Lot or code
    Model No. 989605428532; UDI: (01)00884838073623(21)B305XD, (01)00884838073623(21)F05R7X, (01)00884838073623(21)B1Y35V, (01)00884838073623(21)B305ZX, (01)00884838073623(21)F05K8W, (01)00884838073623(21)B3GWRJ, (01)00884838073623(21)B1Y306, (01)00884838073623(21)B1Y3CG, (01)00884838073623(21)B1Q699, (01)00884838073623(21)B1Y3C4, (01)00884838073623(21)B1RXWK, (01)00884838073623(21)B305X5, (01)00884838073623(21)B305ZK, (01)00884838073623(21)B3NFD5, (01)00884838073623(21)F034C6; Serial No. B305XD, F05R7X, B1Y35V, B305ZX, F05K8W, B3GWRJ, B1Y306, B1Y3CG, B1Q699, B1Y3C4, B1RXWK, B305X5, B305ZK, B3NFD5, F034C6.
    FDA recall number
    Z-2375-2025
    Quantity
    171,322 units
  • L12-5 50 Transducer Probe

    Lot or code
    Model No. 989605418552, 989605408173, 989605438481, 989605420231, 989605412401, 989605408171, 989605362313, 989605396002, 989605387311, 989605412042, 989605418551, 989605409562, 989605412072, 989605409582, 989605440931, 989605361882, 989605440932; UDI: (01)00884838067851(21)B1N28J, (01)00884838067851(21)B1N2GB, (01)00884838067851(21)B20JW8, (01)00884838067851(21)B2TL6N, (01)00884838067851(21)B2TM54, (01)00884838067851(21)B2TNPZ, (01)00884838067851(21)B2TPD0, (01)00884838067851(21)B2ZJYP, (01)00884838067974(21)B24F91, (01)00884838067851(21)B2TMH2, 010088483806785121F0FNVH, 010088483806785121F0FQMW, 010088483806785121F0FQNJ, (01)00884838067851(21)B24NM7, (01)00884838067851(21)B2D9MB, (01)00884838067851(21)B2D9VH, (01)00884838067851(21)B2H8B5, (01)00884838067851(21)B2PL21, (01)00884838067851(21)B40HY0, (01)00884838067851(21)B40J43, (01)00884838067974(21)B1N1NV, (01)00884838067851(21)B3XX7R, (01)00884838067851(21)F088MN, (01)00884838067851(21)B0P7DV, (01)00884838067851(21)B2ZHQ3, (01)00884838067851(21)B40VT9, (01)00884838067851(21)F0589C, (01)00884838067851(21)F0CHBM, (01)00884838067851(21)F0FNT5, (01)00884838067851(21)F0JZNN, 010088483806785121F0FZWN, 010088483806785121F0FZXK, (01)00884838067851(21)B20QHY, (01)00884838067851(21)B20GD4, (01)00884838067851(21)F0JTMB, 010088483806785121B3XJNR, (01)00884838067851(21)B39N0H, (01)00884838067974(21)B0LMXB, (01)00884838067974(21)B2H5LC, (01)00884838067974(21)B2PR9K, (01)00884838067974(21)B2ZK20, (01)00884838067974(21)B2ZW1C,…
    FDA recall number
    Z-2376-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 150,529 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L15-7io Transducer Probe

    Lot or code
    Model No. 989605387852, 989605438421, 989605387341, 989605428532, 989605391083, 989605412042, 989605412082, 989605418963, 989605387851; UDI: (01)00884838061491(21)B2Y4MR, (01)00884838061491(21)B2YR9K, (01)00884838061491(21)F0DWM3, (01)00884838061491(21)F084NQ, (01)00884838061491(21)F0D244, (01)00884838061491(21)F0F19W, (01)00884838061491(21)F0G2RJ, (01)00884838061491(21)F0K6X7, (01)00884838087057(21)B2C14V, (01)00884838061491(21)B1PJC0, (01)00884838061491(21)B24YWK, (01)00884838061491(21)F0L51D, (01)00884838061491(21)F0L51F, (01)00884838061491(21)B31Q0J, (01)00884838061491(21)B3KQTR, (01)00884838061491(21)B2KLR3, (01)00884838061491(21)B2KLZ4, (01)00884838061491(21)B3H0PX, (01)00884838061491(21)B3J6LC, (01)00884838061491(21)F03BYJ, (01)00884838061491(21)F046Q9, (01)00884838061491(21)F04L49, (01)00884838061491(21)F0CXV3, (01)00884838061491(21)F0D5LN, (01)00884838061491(21)B1WGZP, (01)00884838061491(21)F03WVP, (01)00884838061491(21)F044Z4, (01)00884838073623(21)F041TJ, 010088483806149121F025Z4, (01)00884838061491(21)F002TP, (01)00884838061491(21)F06BF7, (01)00884838061491(21)F06C7C, (01)00884838061491(21)F06CWM, (01)00884838061491(21)F06FP7, (01)00884838061491(21)F06GLG, (01)00884838061491(21)F0CZ2Z, (01)00884838061491(21)F0D6R6, 010088483806149121F0G7MD, (01)00884838061491(21)B285T6, (01)00884838061491(21)F00ZL7, (01)00884838061491(21)F0101H, (01)00884838061491(21)F010NB, (01)00884838061491(21)B2QGT2, (01)00884838061491(21)B2X20Q, (01)00884838061491(21)B2YYTP,…
    FDA recall number
    Z-2377-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 185,924 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L15-7IO TRANSDUCER Transducer Probe

    Lot or code
    Model No. N/A; UDI: N/A; Serial No. B1K740.
    FDA recall number
    Z-2378-2025
    Quantity
    171,322 units
  • L17-5 Transducer Probe

    Lot or code
    Model No. 8500-1662-01, 989605366373, 989605388061, 989605366371; UDI: (01)00884838064614(21)B20XYG, (01)00884838064614(21)B2HK04, (01)00884838064614(21)B36C77, (01)00884838064614(21)B22N5X, (01)00884838064614(21)B1YQ5T, (01)00884838064614(21)B30M34, (01)00884838064614(21)B3GX1Q, (01)00884838064614(21)B3H0X1, (01)00884838064614(21)B3P6N2, (01)00884838064614(21)B22N0X, (01)00884838064614(21)B2YF79, (01)00884838064614(21)B1XRFM, (01)00884838064614(21)B2YD0X, (01)00884838064614(21)B3BZNW, (01)00884838064614(21)B2YNKW, (01)00884838064614(21)B1QBQ1, (01)00884838064614(21)B1W401, (01)00884838064614(21)B28MR6, (01)00884838064614(21)B30M32, (01)00884838064614(21)B31CR5, (01)00884838064614(21)B3MJMX, (01)00884838064614(21)B3MK52, (01)00884838064614(21)F093DP, (01)00884838064614(21)B2JDWW, (01)00884838064614(21)B2KN10, (01)00884838064614(21)B1ZR3M, (01)00884838064614(21)B1TVHQ, (01)00884838064614(21)B2NQ1G, (01)00884838064614(21)B1VL0Y, (01)00884838064614(21)B1Q3BM, (01)00884838064614(21)B31VQ3, (01)00884838064614(21)B2LM71, (01)00884838064614(21)F048JT, (01)00884838064614(21)B2ZC0V, (01)00884838064614(21)B2HTPC, (01)00884838064614(21)B326LT, (01)00884838064614(21)B3J3L3, (01)00884838064614(21)F02LCK, (01)00884838064614(21)B285BJ, (01)00884838064614(21)B2G3H4, (01)00884838064614(21)B3GX28, (01)00884838064614(21)B3H0X5, (01)00884838064614(21)B1VL2B, (01)00884838064614(21)B1Q3BL, (01)00884838064614(21)B289YG, (01)00884838064614(21)B1WGX0, (01)00884838064614(21)B1RT7M,…
    FDA recall number
    Z-2379-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 5,807 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L18-5 Transducer Probe

    Lot or code
    Model No. 989605409582, 989605438531; UDI: (01)00884838061590(21)B211XD, (01)00884838061590(21)B29X8C, (01)00884838061590(21)B2VJ24, (01)00884838061590(21)F010Y2, (01)00884838061590(21)F04JYB, (01)00884838061590(21)F06LTM, 010088483806159021F024QP, (01)00884838061590(21)B23T88, (01)00884838061590(21)B2KLW4, (01)00884838061590(21)B2LL93, (01)00884838061590(21)F0DJP0, (01)00884838061590(21)B1ZQTL, (01)00884838061590(21)B2ZCHP, (01)00884838061590(21)B1NXX9, (01)00884838061590(21)B1QBQ0, (01)00884838061590(21)F01CK2, (01)00884838061590(21)B2T427, (01)00884838061590(21)B2X16F, (01)00884838061590(21)F08GP4, (01)00884838061590(21)B1RYJ0, (01)00884838061590(21)B26Q0R, (01)00884838061590(21)B285X3, (01)00884838061590(21)B2860F, (01)00884838061590(21)B286JG, (01)00884838061590(21)B28C28, (01)00884838061590(21)B28MPM, (01)00884838061590(21)B28641, (01)00884838061590(21)F09LKN, (01)00884838061590(21)B1P4DQ, (01)00884838061590(21)B2XFLV, (01)00884838061590(21)F0C64G, (01)00884838061590(21)B1ML3D, (01)00884838061590(21)B1NVNV, (01)00884838061590(21)B3FC9C, (01)00884838061590(21)B3H0RH, (01)00884838061590(21)B3VBR3, (01)00884838061590(21)F00YXV, (01)00884838061590(21)F03DK8, (01)00884838061590(21)F04DBC, (01)00884838061590(21)F04ZZ8, (01)00884838061590(21)F0604G, (01)00884838061590(21)F07YR8, (01)00884838061590(21)F08WT0, (01)00884838061590(21)F098T9, (01)00884838061590(21)F09FJZ, (01)00884838061590(21)F09YZ7, (01)00884838061590(21)F0C6GT, (01)00884838061590(21)F0C6WQ,…
    FDA recall number
    Z-2380-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 44,202 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • L8-4 Transducer Probe

    Lot or code
    Model No. 989605440812, 989605387331, 8500-1659-01; UDI: (01)00884838067684(21)B16YF8, (01)00884838067684(21)B16YCT, (01)00884838067684(21)B1LJXF, (01)00884838067684(21)B1LJXB, (01)00884838067684(21)B0QY2W, (01)00884838067684(21)B0NBTB, (01)00884838067684(21)B2X79D, (01)00884838067684(21)B1M6RD, (01)00884838067684(21)B16YF2, (01)00884838067684(21)B2PFP4, (01)00884838067684(21)B1M6QR, (01)00884838067684(21)B2PFPC, (01)00884838067684(21)B0QY57, (01)00884838067684(21)B36JJ4, (01)00884838067684(21)B11679, (01)00884838067684(21)B0L7X8, (01)00884838067684(21)B1M6QV, (01)00884838067684(21)B11674, (01)00884838067684(21)B0X63X, (01)00884838067684(21)B36JJ5; Serial No. 02RTF9, B16YF8, 02KKZT, B11676, 03H7GL, 02KMJ2, B11673, B16YCT, 037Z0M, B0NBZQ, 02LTKK, 03DTRK, B11672, B1LJXF, B16YCX, B1LJXB, 02KLVP, B0QY2W, 02H2JM, B16YCQ, B0NBTB, 02JCY3, B2X79D, B1M6RD, B16YF2, 038XYV, B04N6N, B2PFP4, 02H34T, B1167B, B0QY7L, B1M6QR, B11675, B16YD1, B2PFPC, B0QY57, B36JJ4, 02LT6G, B11679, B0L7X8, B1168P, 02V8VD, B1M6QV, 02JC66, B11674, 02KKZN, 02V016, B0X63X, B36JJ5, 02NK5J.
    FDA recall number
    Z-2381-2025
    Quantity
    171,322 units
  • L9-3 Transducer Probe

    Lot or code
    Model No. 989605387361, 989605353303, 989605353301, 989605365003; UDI: (01)00884838061767(21)B3PV40, (01)00884838061767(21)B3JRCQ, (01)00884838061767(21)B1ZRXY, (01)00884838061767(21)B1ZRY3, (01)00884838061767(21)B2G5L6, (01)00884838061767(21)B2JFF8, (01)00884838061767(21)B2JZ26, (01)00884838061767(21)B2R3XW, (01)00884838061767(21)B2Y7FN, (01)00884838061767(21)B2YD2P, (01)00884838061767(21)B33XHX, (01)00884838061767(21)B37VC0, (01)00884838061767(21)B2RLWL, (01)00884838061767(21)B3NT2M, (01)00884838061767(21)B2KLVJ, (01)00884838061767(21)B2JF6F, (01)00884838061767(21)B2Y53M, (01)00884838061767(21)B2T51H, (01)00884838061767(21)B3JWZB, (01)00884838061767(21)B3RJ6B, (01)00884838061767(21)B3VBQP, (01)00884838061767(21)B2LL7Z, (01)00884838061767(21)B2Y7BL, (01)00884838061767(21)B22LK7, (01)00884838061767(21)B29JNQ, (01)00884838061767(21)B31CCC, (01)00884838061767(21)B3C6BP, (01)00884838061767(21)B3JRLK, (01)00884838061767(21)B2YNBX, (01)00884838061767(21)B1YQBZ, (01)00884838061767(21)B2ZBZC, (01)00884838061767(21)B1XP34, (01)00884838061767(21)B1Q45M, (01)00884838061767(21)B3J2LX, (01)00884838061767(21)B3RJFP, (01)00884838061767(21)B30GBZ, (01)00884838061767(21)B36PBF, (01)00884838061767(21)F022CP, (01)00884838061767(21)B1NXT9, (01)00884838061767(21)B1QBFT, (01)00884838061767(21)B1TV9K, (01)00884838061767(21)B26N6F, (01)00884838061767(21)B3J3HK, (01)00884838061767(21)F08TJC, (01)00884838061767(21)B251QV, (01)00884838061767(21)B2G38C, (01)00884838061767(21)B2MPDN,…
    FDA recall number
    Z-2382-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 50,362 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • mC12-3 Transducer Probe

    Lot or code
    Model No. 989605495721; UDI: (01)00884838093935(21)F0FKJ7, 010088483809393521F0FRHZ, (01)00884838093935(21)B3XYRY, (01)00884838093935(21)B4161X, (01)00884838093935(21)F07K8J, (01)00884838093935(21)F0JTY0, (01)00884838093935(21)F04YTZ, (01)00884838093935(21)F0550K, (01)00884838093935(21)F094Y9, (01)00884838093935(21)F0GJC9, (01)00884838093935(21)B3XZRM, (01)00884838093935(21)F0KC6Y, (01)00884838093935(21)F0HP38, (01)00884838093935(21)F0F164, (01)00884838093935(21)F0KND1, (01)00884838093935(21)F04YV6, 010088483809393521B3Y0KC, (01)00884838093935(21)F0JZQ6, (01)00884838093935(21)F0JZQ7, (01)00884838093935(21)B3XZ5Q, (01)00884838093935(21)F07WW2, (01)00884838093935(21)F07WW9, (01)00884838093935(21)F07WWF, (01)00884838093935(21)F07WWJ, (01)00884838093935(21)F07WWL, (01)00884838093935(21)F07WWN, (01)00884838093935(21)F07WWP, (01)00884838093935(21)F07WWQ, (01)00884838093935(21)F07ZZ0, (01)00884838093935(21)F081V3, (01)00884838093935(21)F08H3V, (01)00884838093935(21)F05FK6, (01)00884838093935(21)F06GWZ, (01)00884838093935(21)F07D0C, (01)00884838093935(21)F0B94X, (01)00884838093935(21)F0GJCR, (01)00884838093935(21)B3FKZJ, (01)00884838093935(21)B3FKZN, (01)00884838093935(21)F0LC8N, (01)00884838093935(21)F04QBB, (01)00884838093935(21)F04V3M, (01)00884838093935(21)F04V3Q, (01)00884838093935(21)F04XDN, (01)00884838093935(21)F04XF1, (01)00884838093935(21)F04XF6, (01)00884838093935(21)F06XQF, (01)00884838093935(21)F0F4BF, (01)00884838093935(21)F0F4BN, (01)00884838093935(21)F0F7ZJ,…
    FDA recall number
    Z-2383-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 40,154 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • mC7-2 Transducer Probe

    Lot or code
    Model No. 989605486021; UDI: (01)00884838088160(21)F0H6NW, (01)00884838088160(21)B35HXY, (01)00884838088160(21)B3HWCY, (01)00884838088160(21)B3WVGJ, (01)00884838088160(21)B3WW12, (01)00884838088160(21)B3WW13, (01)00884838088160(21)F0JRVZ, (01)00884838088160(21)B40Q3K, (01)00884838088160(21)F04Q9N, (01)00884838088160(21)F09LNY, (01)00884838088160(21)B40GFT, (01)00884838088160(21)B40HCB, (01)00884838088160(21)B3Y5LX, (01)00884838088160(21)F0D2Q3, (01)00884838088160(21)F0F7Z8, (01)00884838088160(21)F0F7Z9, (01)00884838088160(21)F0FGNZ, (01)00884838088160(21)F0FGP7, (01)00884838088160(21)F0F627, (01)00884838088160(21)F04QB2, (01)00884838088160(21)B39RCM, (01)00884838088160(21)B3XFVY, (01)00884838088160(21)B3YRZK, (01)00884838088160(21)B3Y3Z9, (01)00884838088160(21)F0H6P0, (01)00884838088160(21)F0HBW0, (01)00884838088160(21)B35CT2, (01)00884838088160(21)B3HLT9, (01)00884838088160(21)B35CWP, (01)00884838088160(21)B3W4VD, (01)00884838088160(21)B3XWND, (01)00884838088160(21)B3XXY8, (01)00884838088160(21)B3HWNZ, (01)00884838088160(21)B3L4YF, (01)00884838088160(21)B3L583, (01)00884838088160(21)B3PYLN, (01)00884838088160(21)B3TFDC, (01)00884838088160(21)F0B66P, (01)00884838088160(21)F08R96, (01)00884838088160(21)B39R8V, (01)00884838088160(21)B3YXN2, (01)00884838088160(21)F08NCX, (01)00884838088160(21)F0GM7J, 010088483808816021B3XQBK, (01)00884838088160(21)B35CVB, (01)00884838088160(21)B3HM89, (01)00884838088160(21)B2TN5J, (01)00884838088160(21)B32MXT, (01)00884838088160(21)B3L4LN,…
    FDA recall number
    Z-2384-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 35,708 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Mini Multi TEE Transducer Probe

    Lot or code
    Model No. 989605439242, 21381A; UDI: (01)00884838067547(21)B0DTV4, (01)00884838067547(21)B01639, (01)00884838067547(21)B06NT3; Serial No. 0341YT, 0319CQ, 028NZW, B0RF1J, B0F31L, B0DTV4, B0C9XR, 02H0L8, B0KYGV, B01639, B06NT3, 03C68B, 03C8CW.
    FDA recall number
    Z-2385-2025
    Quantity
    171,322 units
  • mL26-8 Transducer Probe

    Lot or code
    Model No. 989606722361; UDI: (01)00884838108592(21)F08WYR, 010088483810859221F0GDHW, 010088483810859221F0GDJ0, (01)00884838108592(21)F0LCBL, (01)00884838108592(21)F0JX0Q, (01)00884838108592(21)F0GT0X, (01)00884838108592(21)F0KVCN, (01)00884838108592(21)F0LC9H, (01)00884838108592(21)F0H4DG, (01)00884838108592(21)F0J4J8, (01)00884838108592(21)F0HX31, (01)00884838108592(21)F0LQC4, (01)00884838108592(21)F0LQBV, (01)00884838108592(21)F0JX1C, (01)00884838108592(21)F0LC9M, (01)00884838108592(21)F0LC9X, (01)00884838108592(21)F0DM3Y, (01)00884838108592(21)F0KC82, (01)00884838108592(21)F0J4J6, (01)00884838108592(21)F0JZ0P, (01)00884838108592(21)F0D1BV, (01)00884838108592(21)F0H2LF, (01)00884838108592(21)F0H2L7, (01)00884838108592(21)F0HGKW, (01)00884838108592(21)F0LGP6, (01)00884838108592(21)F0C91D, (01)00884838108592(21)F0JQV1, (01)00884838108592(21)F0JR9G, (01)00884838108592(21)F0KGDW, (01)00884838108592(21)F0H9FK, (01)00884838108592(21)F0H9G5, (01)00884838108592(21)F0HYV9, (01)00884838108592(21)F0LGPV, (01)00884838108592(21)F0JX0V, 010088483810859221F0G3K8, (01)00884838108592(21)F0JR9T, (01)00884838108592(21)F0JX0H, (01)00884838108592(21)F0JZ0M, (01)00884838108592(21)F0C919, (01)00884838108592(21)F0C9MD, 010088483810859221F0FVL8, (01)00884838108592(21)F0L39V, (01)00884838108592(21)F0LL43, (01)00884838108592(21)F0LGPN, (01)00884838108592(21)F0K1N0, (01)00884838108592(21)F0K705, (01)00884838108592(21)F0K70C, (01)00884838108592(21)F0K70D, (01)00884838108592(21)F0K70G,…
    FDA recall number
    Z-2386-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 18,835 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • OMNI II TEE Transducer Probe

    Lot or code
    Model No. N/A; UDI: N/A; Serial No. 02DBYL, 02P8MV, 038CLG.
    FDA recall number
    Z-2387-2025
    Quantity
    171,322 units
  • OMNI III TEE Transducer Probe

    Lot or code
    Model No. 989605440702, 21378A; UDI: (01)00884838067585(21)B1CW9T, (01)00884838067585(21)B201MC, (01)00884838067585(21)B1CW9Q, (01)00884838067585(21)B1MK5D, (01)00884838067585(21)B1MK5X, (01)00884838067585(21)B198WH, (01)00884838067585(21)B1MKD9, (01)00884838067585(21)B198V3, (01)00884838067585(21)B201JF, (01)00884838067585(21)B201Q0, (01)00884838067585(21)B1991G, (01)00884838067585(21)B17X87, (01)00884838067585(21)B1MKC3, (01)00884838067585(21)B201QF, (01)00884838067585(21)B2FW9F, (01)00884838067585(21)B198VV, (01)00884838067585(21)B07FFB, (01)00884838067585(21)B2FW8Z, (01)00884838067585(21)B201N8, (01)00884838067585(21)B34R1B, (01)00884838067585(21)B05DX2, (01)00884838067585(21)B0LYR8, (01)00884838067585(21)B201K2, (01)00884838067585(21)B198Z2, (01)00884838067585(21)B1991B; Serial No. B1CW9T, 02Y7HM, 02VTGK, B201MC, B1CW9Q, 037CHG, 03G8B1, 03G0BB, B1MK5D, B1MK5X, 02WH6F, 033ZLX, 033W7Z, 03CWJM, 02QK2L, B198WH, 038MJ6, 037F68, B1MKD9, 038GL7, 0344C9, B04LK2, B1991F, B198V3, B1FTDM, 0341PJ, B201JF, 037HLM, 035R8K, B201Q0, B1FTFZ, 02QCRQ, B1991G, 02JL01, B1MK2K, B17X87, 02JJ0T, 031LC8, B1MKC3, B201QF, B1990H, B19918, 03FZXM, 039JWQ, 035PYK, 03436F, B2FW9F, 039PCB, B1MK2F, 031HN7, 02P8GV, B198VV, 02H0WW, 039HQ2, B07FFB, B2FW8Z, 02WJBB, B201N8, 02VP8P, B34R1B, 02YK1Q, B19900, 039Q2L, B05DX2, 02YCFD, 02WCTB, 039PC3, 035V3H, B0LYR8, B201K2, 02PDT7, 032MLN, 035P8R, 038LT2, B0M32G, 03G4GG, B0X7N7, 038HV5, 032MMK, 02QGKV, 028PNR, B198Z2, 032FKW, B198ZV, 03GZ12, 03191G, B1991B,…
    FDA recall number
    Z-2388-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 24 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • S12-4 Transducer Probe

    Lot or code
    Model No. 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361973, 989605361973, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361973, 989605361892, 989605438411, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605438411, 989605361892, 989605361892, 989605361892, 989605361892, 989605361973, 989605361973, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361973, 989605438411, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361973, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605361892, 989605438411, 989605361892, 989605361892,…
    FDA recall number
    Z-2389-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 183,773 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • S3-1 Transducer Probe

    Lot or code
    Model No. 989605387381, 989605361403, 989605440732, 989605361401; UDI: (01)00884838068001(21)B1RDXB, (01)00884838068001(21)B1RQG6, (01)00884838068001(21)B1RMV9, (01)00884838067608(21)B1YRPX, (01)00884838067608(21)B1YT4J, (01)00884838068001(21)B24RFX, (01)00884838068001(21)B1RDX8, (01)00884838068001(21)B20PGC, (01)00884838068001(21)B1RJT1, (01)00884838068001(21)B20PGB, (01)00884838068001(21)B1RDNY, (01)00884838068001(21)B1MW4N, (01)00884838068001(21)B1MVVJ, (01)00884838068001(21)B20PN8, (01)00884838068001(21)B1RDXC; Serial No. 0388D2, B1JWWV, B19RZV, B0RHV4, B0RH23, B1RDXB, B16W47, B1RQG6, B0RHV3, B19QZZ, B19R10, B16JPL, B1RMV9, B19V6V, B1YRPX, B1YT4J, B24RFX, B0RHV1, B0RR1C, B19V02, B10DQ0, B19V08, B1RDX8, B20PGC, B1RJT1, B20PGB, B1JWWZ, B1RDNY, B1MW4N, 03DLPH, B19V05, B19V6N, B1MVVJ, B0RHV5, B0RHV8, B20PN8, B1RDXC, B19V6W, B0RHV2.
    FDA recall number
    Z-2390-2025
    Quantity
    171,322 units
  • S4-1 Transducer Probe

    Lot or code
    Model No. 989605389271, 989605389272, 989605364993; UDI: (01)00884838061415(21)B30VBN, (01)00884838061415(21)F0954W, (01)00884838061415(21)B3WCRT, (01)00884838061415(21)B1WDKX, (01)00884838061415(21)B1WDNQ, (01)00884838064584(21)B13HXX, (01)00884838064584(21)B1MQDX, (01)00884838064584(21)B23LRM, (01)00884838064584(21)B23LRT, (01)00884838061415(21)B1X9Z3, (01)00884838064584(21)B12QQ8, (01)00884838061415(21)B27XB2, (01)00884838061415(21)B307TT, 010088483806141521F01WFB, (01)00884838061415(21)B30THD, (01)00884838061415(21)B1TKB3, (01)00884838061415(21)F08XDL, (01)00884838061415(21)F0B7L5, (01)00884838061415(21)F048QQ, (01)00884838061415(21)F0665R; Serial No. B30VBN, B1GPT4, B1NW75, F0954W, B3WCRT, B1WDKX, B1WDNQ, B13HXX, B1MQDX, B23LRM, B23LRT, B1X9Z3, B12QQ8, B27XB2, B307TT, F01WFB, B1KXRP, B30THD, B1TKB3, F08XDL, F0B7L5, F048QQ, F0665R.
    FDA recall number
    Z-2391-2025
    Quantity
    171,322 units
  • S4-1 Lumify Transducer Probe

    Lot or code
    Model No.989605458222, 989605477101, 989605458221; UDI: (01)00884838085183(21)B3G9Q2, (01)00884838074231(21)B1RCZ0, (01)00884838074231(21)B1RD8W, (01)00884838074231(21)B1RDD9, (01)00884838074231(21)B1W89Q, (01)00884838074231(21)B1W8D2, (01)00884838085183(21)B2D2YM, (01)00884838074231(21)B21YB6, (01)00884838085183(21)B35Y6X, (01)00884838085183(21)B36B15, (01)00884838085183(21)B36B6J, (01)00884838086579(21)B2C7BV, (01)00884838085183(21)F0BH9V, (01)00884838085183(21)F0DDFJ, (01)00884838085183(21)F0DDFL, (01)00884838085183(21)F0DDZD, (01)00884838085183(21)F0DDZF, (01)00884838085183(21)F0DF0H, (01)00884838085183(21)F0FFWW, (01)00884838085183(21)F0FMM0, (01)00884838085183(21)F0FQ22, (01)00884838085183(21)F0FQ79, (01)00884838085183(21)F0FQ7B, (01)00884838085183(21)F089PW, (01)00884838085183(21)B3VTP6, (01)00884838085183(21)B21XV4, (01)00884838085183(21)B2795N, (01)00884838085183(21)B2795P, (01)00884838085183(21)B279F3, (01)00884838085183(21)F03BG3, (01)00884838085183(21)F043BH, (01)00884838085183(21)F0473N, (01)00884838085183(21)F04847, (01)00884838085183(21)F0FKM3, (01)00884838085183(21)F0FL0R, (01)00884838085183(21)F082PN, (01)00884838085183(21)F0KNL7, (01)00884838085183(21)F0KPPW, (01)00884838085183(21)F0KPQ0, (01)00884838085183(21)F051N0, (01)00884838085183(21)F051N3, (01)00884838085183(21)F051N5, (01)00884838085183(21)F056QT, (01)00884838085183(21)F058VC, (01)00884838085183(21)F058VG, (01)00884838085183(21)B1RCKP, (01)00884838085183(21)F03KHH, (01)00884838085183(21)B3D8DD,…
    FDA recall number
    Z-2392-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 196,578 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • S4-2 Transducer Probe

    Lot or code
    Model No. 989605414692, 989605463421, 989605442131, 989605387731, 989605412042, 989605430222, 989605381211, 989605430652, 989605361982, 989605463421; UDI: (01)00884838067844(21)B2TH04, (01)00884838067844(21)B2TH20, (01)00884838067844(21)B2TH5R, (01)00884838067844(21)B2TH94, (01)00884838067844(21)B2TH95, (01)00884838067844(21)B2TRZG, (01)04719873000067(21)F0771Y, (01)04719873000067(21)F0K891, 010471987300006721B24RHN, (01)00884838067844(21)B2ZLWP, (01)00884838067844(21)B1N04C, (01)00884838067844(21)B1N0D2, (01)00884838067844(21)B2PXTV, (01)00884838067844(21)B2PY5Z, (01)00884838067844(21)B2TRRF, (01)00884838067844(21)B1RDM4, (01)04719873000067(21)B3TLRJ, (01)04719873000067(21)B3VZCQ, (01)00884838067844(21)B35KQ5, (01)00884838067844(21)B2ZLK8, (01)00884838067844(21)B2ZTKD, (01)00884838067844(21)B1JVZF, (01)00884838067844(21)F0G5KZ, (01)00884838067844(21)B20GTL, (01)00884838067844(21)B20QB6, 010088483806784421F0B41N, 010088483806784421F0B41W, (01)00884838067844(21)B2PTGM, (01)00884838067844(21)B2TGKT, (01)00884838067844(21)B2TGL2, (01)00884838067844(21)B2TH12, (01)00884838067844(21)B1RF7C, (01)00884838067844(21)B2ZTKF, (01)00884838067844(21)B3W869, (01)00884838067844(21)B3TJP3, (01)00884838067844(21)B3TJWM, 010088483806784421B3Y0V9, (01)00884838067844(21)B2DC5Z, (01)00884838067844(21)B2PN5J, (01)00884838067844(21)F0HWVH, (01)00884838067844(21)F0HYM4, (01)00884838067844(21)F0J4F6, (01)00884838067844(21)F0LH1L, (01)00884838067844(21)B1WVHX, (01)00884838067844(21)B27QW7,…
    FDA recall number
    Z-2393-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 38,215 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • S5-2 Transducer Probe

    Lot or code
    Model No. 989605412082, 989605440682, 989605388091, 21314A, 989605438581, 989605361871, 989605412081, 989605961871, 989605391922, 989605387852, 989605440752, 989605412052, 989605361892, 989605384711, 989605400802; UDI: (01)00884838061637(21)B268VV, (01)00884838061637(21)B2BK4Q, (01)00884838061637(21)B2XV6D, (01)00884838061637(21)B2XVMK, (01)00884838061637(21)B3GLNL, (01)00884838061637(21)B3H17V, (01)00884838061637(21)B3H60B, (01)00884838061637(21)B3RBBZ, (01)00884838061637(21)F00X3V, (01)00884838061637(21)F055Z3, (01)00884838061637(21)F07YX6, (01)00884838061637(21)F0DZGY, (01)00884838061637(21)F0H3M1, (01)00884838061637(21)B3NT5N, (01)00884838061637(21)B1WFML, (01)00884838067561(21)B2BJQ4, (01)00884838067561(21)B2BK64, (01)00884838061637(21)B1Q2GH, (01)00884838061637(21)B1QVY3, 010088483806163721F0G049, (01)00884838061637(21)B3DGJ9, (01)00884838061637(21)F02WHD, (01)00884838061637(21)F0CM58, (01)00884838067561(21)B3D85G, (01)00884838067561(21)F0B3B2, (01)00884838061637(21)B1KYB4, (01)00884838061637(21)B1XMHD, (01)00884838061637(21)B1YT6J, (01)00884838061637(21)B35159, (01)00884838061637(21)F02KM5, (01)00884838061637(21)F0421R, (01)00884838061637(21)F062N7, (01)00884838061637(21)F0D3RL, (01)00884838067561(21)B1837D, (01)00884838067561(21)B19576, (01)00884838061637(21)B2WWLH, (01)00884838061637(21)B3JYHQ, 010088483806163721F02BZK, (01)00884838067561(21)F0918Z, (01)00884838061637(21)B28722, (01)00884838061637(21)F0CWJD, (01)00884838061637(21)F0D9N4, (01)00884838061637(21)B1WFGX,…
    FDA recall number
    Z-2394-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 271,149 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • S7-3t Transducer Probe

    Lot or code
    Model No. 989605406772, 989605439232, 21781A, 989606708921, 989605406771, 989605414122; UDI: (01)00884838061439(21)B21003, (01)00884838061439(21)B3DV68, (01)00884838061439(21)F09Q11, (01)00884838061439(21)B2HXGF, (01)00884838061439(21)F0FJZW, (01)00884838061439(21)F01L0D, (01)00884838061439(21)F02LYB, (01)00884838061439(21)B3JQNC, (01)00884838061439(21)B3JQND, (01)00884838067530(21)B1TDDC, (01)00884838067530(21)B1V2XQ, (01)00884838067530(21)B2HX6Y, N/A, (01)00884838061439(21)F0DZC5, (01)00884838061439(21)B2HXKB, (01)00884838061439(21)B2HXKJ, (01)00884838061439(21)B2HXL1, (01)00884838061439(21)B2HXL4, (01)00884838061439(21)B30FDR, (01)00884838061439(21)B30FDW, (01)00884838061439(21)B30FHX, (01)00884838061439(21)B3DV36, (01)00884838061439(21)F068H7, (01)00884838061439(21)B2HWXC, (01)00884838061439(21)B3JQNB, (01)00884838061439(21)F032XF, N/A, (01)00884838061439(21)B30FFN, (01)00884838061439(21)B3DVDD, 010088483806143921F0GH3M, (01)00884838061439(21)B1X57T, (01)00884838061439(21)B2HWZL, (01)00884838061439(21)B34TFG, (01)00884838061439(21)F080RW, (01)00884838067530(21)B2HX0Q, (01)00884838067530(21)B2HXGM, (01)00884838061439(21)B2HX0H, (01)00884838061439(21)B2HX69, (01)00884838061439(21)F080RX, (01)00884838061439(21)F0DXGZ, (01)00884838067530(21)B0H9H6, (01)00884838067530(21)B0KWX1, (01)00884838067530(21)B22FLD, (01)00884838067530(21)B2HX4M, N/A, N/A, N/A, N/A, (01)00884838061439(21)F0LPPF, (01)00884838061439(21)F0LPPG, (01)00884838061439(21)B1ZJ03, (01)00884838061439(21)B3JQQ5,…
    FDA recall number
    Z-2395-2025
    Quantity
    171,322 units

    The text above is the beginning of the record; 40,466 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Philips Ultrasound, Inc
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

This page lists the first 50 of 69 products of the record. See the full list in the official notice (opens in a new tab). The code check above reads every code of the record.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Philips Ultrasound, Inc, Month: September 2025, Week 2025-W36. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “To provide clarification and labelling to define the useful life of ultrasound transducers in the field.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 97217 on September 3, 2025.

Other notices of Philips Ultrasound, Inc