What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Ultrasound transducer devices were refurbished beyond their useful life.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
385 more codes are in the record below.
Philips 3D9-3v Transducer.
- Lot or code
- Model No. 989605395622, 989605415032, 989605440832, 989605440831, 989605395622 , 989605325181; UDI: (01)00884838061545(21)B0D9FJ, (01)00884838061545(21)B10HWZ, (01)00884838061545(21)B1CQY1, (01)00884838061453(21)B13DR6, (01)00884838061545(21)B1C54Z, (01)00884838061545(21)B1MKLW, (01)00884838061545(21)B1MQGQ, (01)00884838061545(21)B1KWYY, (01)00884838061545(21)B1712M, (01)00884838061545(21)B1M9ZT, (01)00884838061545(21)B1DTRL, (01)00884838061545(21)B1K7BL, (01)00884838061545(21)B1RY9G, (01)00884838061453(21)B0T4K5, (01)00884838061545(21)B0T4K9, (01)00884838061545(21)B0TVWK, (01)00884838061545(21)B0PV0C, (01)00884838061545(21)B1301P, (01)00884838061545(21)B1MQGL, (01)00884838067707(21)036WF9, (01)00884838061545(21)B1LY5T, (01)00884838061545(21)B15F9B, (01)00884838061545(21)B15M6F, (01)00884838061545(21)B0X3K9, (01)00884838061453(21)B0LFYH, (01)00884838061453(21)B0LG5D, (01)00884838061545(21)B1TJB5, (01)00884838061545(21)B18NV8, (01)00884838061545(21)B1F54X, (01)00884838067707(21)B1LY5X, (01)00884838061545(21)B1VR93, (01)00884838061453(21)B0KH69, (01)00884838061545(21)B1VPLF, (01)00884838061545(21)B2BMFY, (01)00884838061545(21)B0R10J, (01)00884838061545(21)B14W77, (01)00884838061545(21)B176PT, (01)00884838061453(21)B0NFZD, (01)00884838061545(21)B0T4K4, (01)00884838061545(21)B0J7V5, (01)00884838061545(21)B0R15J, (01)00884838061545(21)B1GJ9L, (01)00884838061545(21)B0WH4P, (01)00884838061453(21)B0LWCV, (01)00884838067707(21)03BMY9, (01)00884838067707(21)B05DJQ,…
- FDA recall number
- Z-2048-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 2,633 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips C5-1 Transducer.
- Lot or code
- Model No.989605412042, 989605438541, 989605365001, 989605365003; UDI: (01)00884838061613(21)B1M9V1, (01)00884838061613(21)B13H1L; Serial No. B1M9V1, B142ZP, B13H1L, B1LGKZ, B0T0M5, 03DBJJ, 036JBL, B01RWL, B01MZT, 0393MP, 03DWZC, 03FVH9, B02T1M, 03BM2V, B0FN6H, B08TM5, 039D7K, B00RZ7, 03BW39, 0397YZ, B0HG6F, 03937G, 03D7JL, 03C9YZ, 039CNL, 03DVPT, 037VBM, 03DYMH, 03BM6P, B0D9HC, 03BXT8, 03BLG3.
- FDA recall number
- Z-2049-2025
- Quantity
- 5,230 units
Philips C5-2 Transducer.
- Lot or code
- Model No.989605389242, 989605454101, 989605398211, 989605450353, 989605451181, 989605477131, 989605450351, 989605450352; UDI: (01)00884838067424(21)B36LWB, (01)00884838067424(21)B2YZPM, (01)00884838067424(21)B34JQB, (01)00884838067424(21)F01L9N, (01)00884838067424(21)B3V8CH, (01)00884838067424(21)B3LJYB, (01)00884838067424(21)B2C4FQ, (01)00884838067424(21)F010MV, (01)00884838067424(21)B3VGXG, (01)00884838067424(21)F03968, (01)00884838067424(21)B31MT5, (01)00884838067424(21)B28T00, (01)00884838067424(21)B2C44Y, (01)00884838067424(21)B3D56M, (01)00884838067424(21)B22P69, (01)00884838067424(21)B3Q81Z, (01)00884838067424(21)B20145, (01)00884838067424(21)B3R9R3, (01)00884838067424(21)B3G9N5, (01)00884838067424(21)F0174Y, (01)00884838067424(21)B200TW, (01)00884838067424(21)B224WW, (01)00884838067424(21)B3C3W7, (01)00884838067424(21)B2JX46, (01)00884838067424(21)B2VM4B, (01)00884838067424(21)B30VBF, (01)00884838067424(21)B3WD3C, (01)00884838067424(21)B39XCG, (01)00884838067424(21)B2L1X5, (01)00884838067424(21)B2VLYK, (01)00884838067424(21)B3LZ50, (01)00884838067424(21)B3V9LJ, (01)00884838067424(21)B25N1F, (01)00884838067424(21)B231L5, (01)00884838067424(21)B31MT1, (01)00884838067424(21)B2C3W9, (01)00884838067424(21)F016D1, (01)00884838067424(21)B22NVD, (01)00884838067424(21)F008M1, (01)00884838067424(21)B1XBNC, (01)00884838067424(21)B30V7Y, (01)00884838067424(21)B34JXT, (01)00884838067424(21)B3QQX0, (01)00884838067424(21)B2JWTF, (01)00884838067424(21)B1Y1D9,…
- FDA recall number
- Z-2050-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 11,229 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips C9-5EC Transducer.
- Lot or code
- Model No.989605440801; UDI: N/A; Serial No. 03307Y, 02KMNY, 02JC8C, 02Q6LP.
- FDA recall number
- Z-2051-2025
- Quantity
- 5,230 units
Philips eL18-4 Transducer.
- Lot or code
- Model No.989605456291, 989605431531; UDI: (01)00884838070813(21)B2JXFM, (01)00884838070813(21)B25L9M, (01)00884838070813(21)B25LJ0, (01)00884838070813(21)B25LY7, (01)00884838070806(21)B29W04; Serial No. B2JXFM, B25L9M, B28MJZ, B25LJ0, B25LY7, B29W04.
- FDA recall number
- Z-2052-2025
- Quantity
- 5,230 units
Philips eL18-4 EMT Transducer.
- Lot or code
- Model No.989605431531; UDI: (01)00884838070806(21)B29W3Y; Serial No. B29W3Y.
- FDA recall number
- Z-2053-2025
- Quantity
- 5,230 units
Philips L12-3 Transducer.
- Lot or code
- Model No.989605412072, 989605412071, 989605438571, 989605348241; UDI: (01)00884838061620(21)B0TLJT, (01)00884838061620(21)B10GRD, (01)00884838061620(21)B0PYG5, (01)00884838061620(21)B0HF63, (01)00884838061620(21)B0WMR0, (01)00884838061620(21)B15HGH, (01)00884838061620(21)B0Q3CK, (01)00884838061620(21)B0K211, (01)00884838061620(21)B138Y9, (01)00884838061620(21)B11C62, (01)00884838061620(21)B0NK8H, (01)00884838061620(21)B1396N, (01)00884838061620(21)B139FV, (01)00884838061620(21)B0PY5N, (01)00884838061620(21)B16XQR, (01)00884838061620(21)B126FZ, (01)00884838061620(21)B18R5Q, (01)00884838061620(21)B0QKNF, (01)00884838061620(21)B0R3XG, (01)00884838061620(21)B0QN7N, (01)00884838061620(21)B14ZY4, (01)00884838061620(21)B18R9K, (01)00884838061620(21)B1500H, (01)00884838061620(21)B1KB6Q, (01)00884838061620(21)B15NHL, (01)00884838061620(21)B170P0, (01)00884838061620(21)B1CPKD, (01)00884838061620(21)B0TKYJ, (01)00884838061620(21)B144FW, (01)00884838061620(21)B10QGD, (01)00884838061620(21)B13W02, (01)00884838061620(21)B126J4, (01)00884838061620(21)B14ZY0, (01)00884838061620(21)B1B5CB, (01)00884838061620(21)B0PYKQ, (01)00884838061620(21)B1507L, (01)00884838061620(21)B0NKBB, (01)00884838061620(21)B0QN14, (01)00884838061620(21)B10H8J, (01)00884838061620(21)B1410P, (01)00884838061620(21)B12N2J, (01)00884838061620(21)B12P39, (01)00884838061620(21)B16XG6, (01)00884838061620(21)B1503X, (01)00884838061620(21)B0QVTZ, (01)00884838061620(21)B11C3C, (01)00884838061620(21)B15NHC,…
- FDA recall number
- Z-2054-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 2,586 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips L12-4 Transducer.
- Lot or code
- Model No. 989605427122, 989605451171, 989605434813, 989605434812, 989605477121, 989605434811; UDI: (01)00884838072305(21)B1QJ17, (01)00884838061644(21)B2BBRN, (01)00884838061644(21)B1GL69, (01)00884838072305(21)B1YHN7, (01)00884838061644(21)B186QV, (01)00884838072305(21)B1QJ0G, (01)00884838072305(21)B21GPF, (01)00884838072305(21)B1QJ27, (01)00884838072305(21)B1QJ0V, (01)00884838061644(21)B14YT4, 010088483808516921B1L72F, (01)00884838072305(21)B21GNP, (01)00884838072305(21)B21GNR, (01)00884838072305(21)B1YHPV, (01)00884838072305(21)B21GRD, (01)00884838072305(21)B21GMK, (01)00884838085169(21)B1KFNJ, (01)00884838085169(21)B1QGTW, (01)00884838085169(21)B1L78H, (01)00884838072305(21)B1YHNZ, (01)00884838072305(21)B21GQ1, (01)00884838072305(21)B36LQZ, 010088483807230521B21GP4, (01)00884838072305(21)B1YHNK, 010088483808516921B27DYL, (01)00884838085169(21)B1XFGG, (01)00884838072305(21)B21GQH, (01)00884838072305(21)B329WD, (01)00884838085169(21)B1QGXP, (01)00884838072305(21)B21GQ6, (01)00884838072305(21)B21GR5, (01)00884838085169(21)B1KFG2, (01)00884838085169(21)B1QH54, 010088483808516921B1KFMW, (01)00884838085169(21)B1L75L, (01)00884838085169(21)B1QH17, (01)00884838085169(21)B1L762, (01)00884838085169(21)B1JNQF, (01)00884838085169(21)B1QGYV, (01)00884838085169(21)B1L744, (01)00884838085169(21)B21ZCX, (01)00884838085169(21)B1KFL7, (01)00884838085169(21)B1L75N, (01)00884838085169(21)B23HB3, (01)00884838085169(21)B1L7BD, (01)00884838085169(21)B2Y9FR, (01)00884838072305(21)B21GQV,…
- FDA recall number
- Z-2055-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 8,207 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips L8-4 Transducer.
- Lot or code
- Model No. 989605440811, 989605440812; UDI: (01)00884838067684(21)B0LZNN, (01)00884838067684(21)03BPVM, (01)00884838067684(21)02KMFZ, (01)00884838067684(21)02HJDK, (01)00884838067684(21)02FWNQ, (01)00884838067684(21)02KKZT, (01)00884838067684(21)03H7GL, (01)00884838067684(21)02LTKK, (01)00884838067684(21)02H2JM, (01)00884838067684(21)02KKVB, (01)00884838067684(21)038XYV, (01)00884838067684(21)02DLVR, (01)00884838067684(21)02CCM5, (01)00884838067684(21)02LT6G, (01)00884838067684(21)02KMKK, (01)00884838067684(21)02V8VD, (01)00884838067684(21)037Z2B, (01)00884838067684(21)02NK5J; Serial No. B0LZNN, 03BPVM, 02KMFZ, 02HJDK, 02KLBP, 02FWNQ, 02RTF9, 02KKZT, 03H7GL, 02LTKK, 02H2JM, 02KKVB, 038XYV, 02H34T, 02DLVR, 02CCM5, 02LT6G, 02KMKK, 02V8VD, 02JC66, 02V016, 037Z2B, 02NK5J.
- FDA recall number
- Z-2056-2025
- Quantity
- 5,230 units
Philips L9-3 Transducer.
- Lot or code
- Model No. 989605353301, 989605353303, 989605387361; UDI: N/A; Serial No. B0ZTD1, 03F40C, 03923H, 0314BQ, 03712H, 033HGC, 03CJ9M, 035JY0, 03CFTH, 0399XP, 030FNL, 037R81, 02ZH61, 0358PF, 033JGV, 03FN82, 03230B, 03145Y, 02XYB3, 02XWH4, 02Z1D9, 03591C, 0359Q1, 03598Y, 035JVM, 02WVF0, 02WYT1, 03CFPG, 02XWN1, 02ZJR6, 02NV7G, 0305WC, 036XY3, 02XY9D, 03FNPD, 02ZJPH, B08PVN, 0373HX, B0NP6L, 03F364, 03922W, 02X173, 02WYQX, 02ND56, 0367GW, 0315P3, 02QMTL, 038W72, 03315L, 0358PV, 033JK7, 03128V, 032297.
- FDA recall number
- Z-2057-2025
- Quantity
- 5,230 units
Philips S5-1 Transducer.
- Lot or code
- Model No. 989605412082, 989605440682, 989605440681, 989803002101; UDI: (01)00884838061637(21)B1FTT9, (01)00884838061637(21)B18351, (01)00884838061637(21)B1F61P, (01)00884838061637(21)B13G8Q, (01)00884838061637(21)B1F8HC, (01)00884838061637(21)B13G72, (01)00884838061637(21)B15D2R, (01)00884838061637(21)B0R53Y, (01)00884838061637(21)B15CCP, (01)00884838061637(21)B194WG, (01)00884838061637(21)B0ZZF8, (01)00884838061637(21)B15NQ6, (01)00884838061637(21)B170FW, (01)00884838061637(21)B15P4B, (01)00884838061637(21)B0QNTP, (01)00884838061637(21)B140K4, (01)00884838061637(21)B10JZD, (01)00884838061637(21)B15P33, (01)00884838061637(21)B1BGV2, (01)00884838061637(21)B1951N, (01)00884838061637(21)B0R4FP, (01)00884838061637(21)B0R534, (01)00884838061637(21)B10QPH, (01)00884838061637(21)B1PBWF, (01)00884838061637(21)B15D10, (01)00884838061637(21)B16X3N, (01)00884838061637(21)B18VVZ, (01)00884838061637(21)B1L28V, (01)00884838061637(21)B1KY4P, (01)00884838061637(21)B0XB7F, (01)00884838061637(21)B0R40C, (01)00884838061637(21)B140T0, (01)00884838061637(21)B1KY75, (01)00884838061637(21)B14094, (01)00884838061637(21)B194ZQ, (01)00884838061637(21)B15CZZ, (01)00884838061637(21)B0XB6V, (01)00884838061637(21)B1653N, (01)00884838061637(21)B10K9X, (01)00884838061637(21)B12MKD, (01)00884838061637(21)B12VC3, (01)00884838061637(21)B0ZZFJ, (01)00884838061637(21)B15D28, (01)00884838061637(21)B14083, (01)00884838061637(21)B1BGTY, (01)00884838061637(21)B10K8R, (01)00884838061637(21)B1F60B,…
- FDA recall number
- Z-2058-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 23,255 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips S7-2 Transducer.
- Lot or code
- Model No. 989605440772, 989605440771, 989803002261; UDI: (01)00884838067646(21)032MLP, (01)00884838067646(21)B1MMJ8, (01)00884838067646(21)02M708, (01)00884838067646(21)02Q7XJ, (01)00884838067646(21)03CX7R, (01)00884838067646(21)031FV7, (01)00884838067646(21)03C7PW, (01)00884838067646(21)02MJXP, (01)00884838067646(21)02MD57, (01)00884838067646(21)B07FCJ, (01)00884838067646(21)037HLQ, (01)00884838067646(21)B0D3FH, (01)00884838067646(21)031CCH, (01)00884838067646(21)B0M2ZJ, (01)00884838067646(21)039TX5, (01)00884838067646(21)02P5NP, (01)00884838067646(21)031KD1, (01)00884838067646(21)039J9F, (01)00884838067646(21)B0X1K1, (01)00884838067646(21)B0X19Q, (01)00884838067646(21)B04LKL, (01)00884838067646(21)B02VFV, (01)00884838067646(21)03C76J, (01)00884838067646(21)0377MN, (01)00884838067646(21)02Q81T, (01)00884838067646(21)032KCR, (01)00884838067646(21)037HKZ, (01)00884838067646(21)02YH4N, (01)00884838067646(21)02ZTV7, (01)00884838067646(21)B03N8W, (01)00884838067646(21)B0JF6L, (01)00884838067646(21)02VPXJ, (01)00884838067646(21)02MJ8V, (01)00884838067646(21)039GDN, (01)00884838067646(21)B02VM7, (01)00884838067646(21)03G2YM, (01)00884838067646(21)03C229, (01)00884838067646(21)03408Y, (01)00884838067646(21)B14389, (01)00884838067646(21)037FG5, (01)00884838067646(21)03C22H, (01)00884838067646(21)0360L6, (01)00884838067646(21)02VYX4, (01)00884838067646(21)038ND8, (01)00884838067646(21)03CTXP, (01)00884838067646(21)038HV0, (01)00884838067646(21)03H8P2, (01)00884838067646(21)B0D3CJ,…
- FDA recall number
- Z-2059-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 3,273 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips S7-3t Transducer.
- Lot or code
- Model No. 989605440772, 989605440771, 989605439232, 989605406772, 989605439231, 989606708921, 989803002661; UDI: (01)00884838061439(21)B2103Y, (01)00884838061439(21)B1J2M7, (01)00884838061439(21)B2HX4N, (01)00884838061439(21)B187KZ, (01)00884838061439(21)B2HXHR, (01)00884838061439(21)B2104Q, (01)00884838061439(21)B178YZ, (01)00884838061439(21)B22VWR, (01)00884838061439(21)B37JGY, (01)00884838061439(21)B131BH, (01)00884838061439(21)B3DVFC, (01)00884838061439(21)B1BX6R, (01)00884838061439(21)B3JQF5, (01)00884838061439(21)B1GVLN, (01)00884838061439(21)B1WBJZ, (01)00884838061439(21)B3QBXJ, (01)00884838061439(21)B1ZKPV, (01)00884838061439(21)F01W1D, (01)00884838061439(21)B34PDQ, (01)00884838061439(21)B2HX16, (01)00884838061439(21)B3JQFH, (01)00884838061439(21)B3JQQ4, (01)00884838061439(21)B3DV7C, (01)00884838061439(21)B2HWZV, (01)00884838061439(21)B3DV6Y, (01)00884838061439(21)B37JJC, (01)00884838061439(21)B210C8, (01)00884838061439(21)B2HXBW, (01)00884838061439(21)B1FKLQ, (01)00884838061439(21)F001JV, (01)00884838061439(21)B2HX6C, (01)00884838061439(21)B18C91, (01)00884838061439(21)B3DV8Q, (01)00884838061439(21)B3JQFP, (01)00884838061439(21)B3DVCK, (01)00884838061439(21)B3DVCH, (01)00884838061439(21)B209GT, (01)00884838061439(21)B2106D, (01)00884838061439(21)B2HX29, (01)00884838061439(21)B3DVDC, (01)00884838061439(21)F03TQ7, (01)00884838061439(21)B1HHV9, (01)00884838061439(21)B0PK3Y, (01)00884838061439(21)B3JQQ0, (01)00884838061439(21)B3QBWT, (01)00884838061439(21)F001JQ,…
- FDA recall number
- Z-2060-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 15,339 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips S8-3t Transducer.
- Lot or code
- Model No. 989605420183, 989605431171, 989605439232, 989605406772, 989605439231, 989606708921, 989803002661; UDI: (01)00884838073524(21)B2HWZ5, (01)00884838073524(21)B34PCJ, (01)00884838073524(21)B34PBW, (01)00884838073524(21)B37JHN, (01)00884838073524(21)B34P9K, (01)00884838073524(21)B2106G, (01)00884838073524(21)B210DC, (01)00884838073524(21)B2101L, (01)00884838073524(21)B2HX89, (01)00884838073524(21)B2HX1G, (01)00884838073524(21)B2HX7N, (01)00884838073524(21)B2HXBB, (01)00884838073524(21)B2HWZ6, (01)00884838073524(21)B2HXB8, (01)00884838073524(21)B2HX1D, (01)00884838073524(21)B2HWXV, (01)00884838073524(21)B209LJ, (01)00884838073524(21)B37JHP, (01)00884838073524(21)B2HWXW, (01)00884838073524(21)B2105P, (01)00884838073524(21)B1Z0Z8, (01)00884838073524(21)B210D8, (01)00884838073524(21)B21004, (01)00884838073524(21)B1Z0ZD, (01)00884838073524(21)B2HXDN, (01)00884838067523(21)B0XTVJ, (01)00884838073524(21)B2HWXX, (01)00884838073524(21)B20ZYV, (01)00884838073524(21)B2HWWW, (01)00884838073524(21)B2HBQR, (01)00884838073524(21)B2HXK7, (01)00884838073524(21)B2109B, (01)00884838073524(21)B2HXDQ, (01)00884838073524(21)B2109V, (01)00884838073524(21)B34PGV, (01)00884838073524(21)B2107T, (01)00884838073524(21)B34PGQ, (01)00884838073524(21)B3DVD7, (01)00884838073524(21)B34PB8, (01)00884838073524(21)B3DV8N, (01)00884838073524(21)B2100W, (01)00884838073524(21)B21047, (01)00884838073524(21)B2109H, (01)00884838073524(21)B2HXG0, (01)00884838073524(21)B2HX3K, (01)00884838073524(21)B209LK,…
- FDA recall number
- Z-2061-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 7,569 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips V6-2 Transducer.
- Lot or code
- Model No. 989605357563, 989605415042, 989605437982, 989605387201, 989605357561; UDI: (01)00884838061651(21)B1F8BF, (01)00884838061798(21)036HF1, (01)00884838061651(21)B1MB1Y, (01)00884838061651(21)B1P7X7, (01)00884838061651(21)B20WDH, (01)00884838061460(21)B1CYQ2, (01)00884838061460(21)B0ZP1G, (01)00884838061460(21)B0GX06, (01)00884838061460(21)B0T41B, (01)00884838061651(21)B1RYY9, (01)00884838061651(21)B20X2D, (01)00884838061460(21)B1K7CX, (01)00884838061798(21)036GQH, (01)00884838061798(21)036QQF, (01)00884838061798(21)B0K8G6, (01)00884838061798(21)B0NLYT, (01)00884838061798(21)03GKBK, (01)00884838061798(21)03CDKW, (01)00884838061798(21)031PP2, (01)00884838061798(21)B0JJ3P, (01)00884838061798(21)B09HK1, (01)00884838061798(21)0397KL, (01)00884838061798(21)036ZPK, (01)00884838061798(21)036QR2, (01)00884838061798(21)02Y05W, (01)00884838061798(21)03BVQR, (01)00884838061798(21)02WV2J, (01)00884838061798(21)0305DR, (01)00884838061798(21)03933Q, (01)00884838061798(21)038TQR, (01)00884838061798(21)02W80R, (01)00884838061798(21)B0GYV9, (01)00884838061798(21)03H3XM, (01)00884838061798(21)03CCG5, (01)00884838061798(21)031T4C, (01)00884838061798(21)036YYJ, (01)00884838061651(21)B0K8C4, (01)00884838061798(21)036H1C, (01)00884838061798(21)0365ZV, (01)00884838061798(21)036WXV, (01)00884838061798(21)02Z1TZ, (01)00884838061798(21)032C2R, (01)00884838061798(21)031V8K, (01)00884838061798(21)03CBP2, (01)00884838061798(21)038WHD, (01)00884838061798(21)036ZPJ, (01)00884838061798(21)031N36,…
- FDA recall number
- Z-2062-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 4,409 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips VL13-5 Transducer.
- Lot or code
- Model No. 989605409592, 989605372211, 989605372213; UDI: (01)00884838061606(21)B1C23Z, (01)00884838061606(21)B252D4, (01)00884838061606(21)B12ZWV, (01)00884838061606(21)B12QQC, (01)00884838061606(21)B23P58, (01)00884838061606(21)B0X3RL, (01)00884838061606(21)B12RN0, (01)00884838061606(21)B13BCL, (01)00884838061606(21)B1F233, (01)00884838061606(21)B2DWNC, (01)00884838061606(21)B12RN4, (01)00884838061606(21)B12ZTJ, (01)00884838061606(21)B1XR83, (01)00884838061606(21)B13DFT, (01)00884838061606(21)B12QVB, (01)00884838061606(21)B18LZX, (01)00884838067493(21)03FNNG, (01)00884838067493(21)038VY0, (01)00884838067493(21)03926P, (01)00884838067493(21)0367GM, (01)00884838067493(21)036BRP; Serial No. B1C23Z, B252D4, B12ZWV, B12QQC, B23P58, B0X3RL, B12RN0, B13BCL, B1F233, B2DWNC, B12RN4, B12ZTJ, B1XR83, B13DFT, B12QVB, B18LZX, 03FNNG, 038VY0, 03926P, 0367GM, 0399WM, 036BRP, 039250.
- FDA recall number
- Z-2063-2025
- Quantity
- 5,230 units
Philips X5-1 Transducer.
- Lot or code
- Model No. 989605400802; UDI: (01)00884838061422(21)B2XMDQ, (01)00884838061422(21)B3H27D, (01)00884838061422(21)B2ZXZB, (01)00884838061422(21)B26JC4, (01)00884838061422(21)B3GYXN, (01)00884838061422(21)B2ZXD8, (01)00884838061422(21)B3GJB2, (01)00884838061422(21)B2MQZ3, (01)00884838061422(21)B3L198, (01)00884838061422(21)B3KC2D, (01)00884838061422(21)B3MTN1, (01)00884838061422(21)B2894C; Serial No. B2XMDQ, B3H27D, B2ZXZB, B26JC4, B3GYXN, B2ZXD8, B1CGYC, B3GJB2, B2MQZ3, B3L198, B3KC2D, B3MTN1, B2894C, B1FHTZ.
- FDA recall number
- Z-2064-2025
- Quantity
- 5,230 units
Philips X6-1 Transducer.
- Lot or code
- Model No. 989605409282, 989605438511, 989605357143, 989605357141; UDI: (01)00884838061446(21)B1HT1C, (01)00884838061446(21)B1HT47, (01)00884838061446(21)B1W368, (01)00884838061446(21)B1WBDL, (01)00884838061446(21)B1GTFL, (01)00884838061446(21)B11YW3, (01)00884838061446(21)B15QX1, (01)00884838061446(21)B14LVN, (01)00884838061446(21)B22R48, (01)00884838061446(21)B11XQP, (01)00884838061446(21)B1HTJ3, (01)00884838061446(21)B11Z3B, (01)00884838061446(21)B0WX2W, (01)00884838061446(21)B0WYZ1, (01)00884838061446(21)B131LH, (01)00884838061446(21)B1C1C3, (01)00884838061446(21)B1JBBF, (01)00884838061446(21)B101WK, (01)00884838061446(21)B14M20, (01)00884838061446(21)B1GVWH, (01)00884838061446(21)B1H6V3, (01)00884838061446(21)B101TF, (01)00884838061446(21)B1JCYV, (01)00884838061446(21)B2HQD0, (01)00884838061446(21)B27YP9, (01)00884838061446(21)B21DZ3, (01)00884838061446(21)B0Z7G7, (01)00884838061446(21)B0V082, (01)00884838061446(21)B22QVL, (01)00884838061446(21)B14MMW, (01)00884838061446(21)B24VGG, (01)00884838061446(21)B21CGY, (01)00884838061446(21)B1GTDB, (01)00884838061781(21)B157R5, (01)00884838061781(21)B1903C, (01)00884838061781(21)B187M7, (01)00884838061446(21)B21CH0, (01)00884838061781(21)B14MMX, (01)00884838061446(21)B2HQCW, (01)00884838061446(21)B18YDK, (01)00884838061781(21)B13R4G, (01)00884838061781(21)B11X02, (01)00884838061446(21)B18Z17, (01)00884838061446(21)B0V150, 010088483806144621B1WQHH; Serial No. B1HT1C, B1HT47, B1W368, B1WBDL, B1GTFL, B11YW3, B15QX1, B14LVN, B22R48,…
- FDA recall number
- Z-2065-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 384 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips X7-2t Transducer.
- Lot or code
- Model No. 989605350454, 989605361911, 989605414122, 989605461941, 989605350452, 989605350451, 989605455171, 989605350453, 989605438611; UDI: Unique Device Identifier (UDI), (01)00884838061668(21)B144RL, (01)00884838061668(21)B28L3F, (01)00884838061668(21)B26MZ9, (01)00884838061668(21)B3DT21, (01)00884838061668(21)B3DP0K, (01)00884838061668(21)B2JJFV, (01)00884838061668(21)B0T9J4, (01)00884838061668(21)B2XCH6, (01)00884838061668(21)B3DRBG, (01)00884838061668(21)B21DC0, (01)00884838061668(21)B21C1Q, (01)00884838061668(21)B13R5Z, (01)00884838061668(21)B2JJM6, (01)00884838061668(21)B2JJ71, (01)00884838061668(21)B1ZGT3, (01)00884838061668(21)B1C112, (01)00884838061668(21)B1JMPG, (01)00884838061668(21)B1ZH27, (01)00884838061668(21)B1XZ2L, (01)00884838061668(21)B3DT3K, (01)00884838061668(21)B20TVK, (01)00884838061668(21)B1BDJG, (01)00884838061668(21)B1TY40, (01)00884838061668(21)B1ZGM3, (01)00884838061668(21)B1J660, (01)00884838061668(21)B1TCB5, (01)00884838061668(21)B238J6, (01)00884838061668(21)B2VQN4, (01)00884838061668(21)B1ZH4K, (01)00884838061668(21)B191TJ, (01)00884838061668(21)B21DL6, (01)00884838061668(21)B1TB82, (01)00884838061668(21)B23F6J, (01)00884838061668(21)B15PQH, (01)00884838061668(21)B22VHF, (01)00884838061668(21)B110QK, (01)00884838061668(21)B1W3GK, (01)00884838061668(21)B1HCHJ, (01)00884838061668(21)B1ZGYV, (01)00884838061668(21)B3M1H8, (01)00884838061668(21)B1W32L, (01)00884838061668(21)B18B6P, (01)00884838061668(21)B2VQ9X, (01)00884838061668(21)B20TTV,…
- FDA recall number
- Z-2066-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 66,871 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips X8-2t Transducer.
- Lot or code
- Model No. 989605455171; UDI: (01)00884838069718(21)B2JH74, (01)00884838069718(21)B29746, (01)00884838069718(21)B1ZGLZ, (01)00884838069718(21)B34TNV, (01)00884838069718(21)B2JJHM, (01)00884838069718(21)B34TLN, (01)00884838069718(21)B2C84F, (01)00884838069718(21)B2FTXW, (01)00884838069718(21)B24WXT, (01)00884838069718(21)B2T8C5, (01)00884838069718(21)B2JJGX, (01)00884838069718(21)B2T86R, (01)00884838069718(21)B2C8L5, (01)00884838069718(21)B38VJZ, (01)00884838069718(21)B2JGTC, (01)00884838069718(21)B2VQFK, (01)00884838069718(21)B2XD5P, (01)00884838069718(21)B2B8WQ, (01)00884838069718(21)B2JGNL, (01)00884838069718(21)B274VJ, (01)00884838069718(21)B2BN6Q, (01)00884838069718(21)B1ZGLQ, (01)00884838069718(21)B2NMKC, (01)00884838069718(21)B2JJD0, (01)00884838069718(21)B24WRV, (01)00884838069718(21)B39B8M, (01)00884838069718(21)B34Z1Y, (01)00884838061668(21)B1ZH9T, (01)00884838069718(21)B2VQNN, (01)00884838069718(21)B2FVW4, (01)00884838069718(21)B241C7, (01)00884838069718(21)B2JJPB, (01)00884838069718(21)B28JKF, (01)00884838069718(21)B2T85C, (01)00884838069718(21)B2FDPK, (01)00884838069718(21)B2JHK4, (01)00884838069718(21)B2GCL7, (01)00884838069718(21)B2JJBM, (01)00884838069718(21)B2XDJW, (01)00884838069718(21)B3NCQK, (01)00884838069718(21)B3CN4X, (01)00884838069718(21)B2NMF2, (01)00884838069718(21)B2JH9N, (01)00884838069718(21)B2FT86, (01)00884838069718(21)B2CGH0, (01)00884838069718(21)B25DGP, (01)00884838069718(21)B34Z7H, (01)00884838069718(21)B2B7HL, (01)00884838069718(21)B29MMT,…
- FDA recall number
- Z-2067-2025
- Quantity
- 5,230 units
The text above is the beginning of the record; 22,253 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Philips XL14-3 Transducer.
- Lot or code
- Model No. 989605409251; UDI: (01)00884838074613(21)B1ZL08; Serial No. B1ZL08.
- FDA recall number
- Z-2068-2025
- Quantity
- 5,230 units
Distribution (FDA): Domestic: Nationwide Distribution; Foreign: Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nicaragua, Oman, Philippines, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, United Arab Emirates, Venezuela, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips Ultrasound, Inc
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.