Why it was recalled
Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R
- Lot or code
- REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28
- FDA recall number
- Z-2175-2025
- Quantity
- 55
Distribution (FDA): Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Remote Diagnostic Technologies Ltd.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.