Remote Diagnostic Technologies Ltd. recalls
5 notices naming Remote Diagnostic Technologies Ltd. as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Remote Diagnostic Technologies Ltd.: Philips Tempus Pro Patient Monitor
Hazard (FDA): Patient monitor measures Intracranial Pressure (ICP) and Bladder Pressure (BDR) but ICP and BDR measurements have not been tested or validated for accuracy. If healthcare providers base clinical judgment on unvalidated…
FDAMedical devices
Remote Diagnostic Technologies Ltd.: Inseego USB8 4G Dongle Kit
Hazard (FDA): Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead…
FDAMedical devices
Remote Diagnostic Technologies Ltd.: Philips Tempus Pro Patient Monitor
Hazard (FDA): Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May…
FDAMedical devices
Remote Diagnostic Technologies Ltd.: Tempus Pro Patient Monitor
Hazard (FDA): software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
FDAMedical devices
Remote Diagnostic Technologies Ltd.: Tempus LS-Manual Defibrillator Model 00-3020 The Tempus LS-Manual is a defibrillator…
Hazard (FDA): Due to incorrect labels applied to devices during servicing. Labels contain the incorrect product name and CAT number.
FDAMedical devices