Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 96107: Sysmex America, Inc.: Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6.

FDA recall 96107 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

26 more codes are in the record below.

  • Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

    Lot or code
    UDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967
    FDA recall number
    Z-1113-2025
    Quantity
    32 units

Distribution (FDA): US Nationwide.

Source

Issued by
U.S. Food and Drug Administration
Firm
Sysmex America, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Sysmex America, Inc., Month: February 2025, Week 2025-W07. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.