Sysmex America, Inc. recalls
2 notices naming Sysmex America, Inc. as the recalling firm, since February 12, 2025. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Sysmex America, Inc.: TS-10/TS-10H Tube Sorter
Hazard (FDA): Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
FDAMedical devices
Sysmex America, Inc.: Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6.
Hazard (FDA): Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
FDAMedical devices