What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.
- Lot or code
- UDI-DI 00827002246638, Lot numbers W4844245, exp. 6/10/2027; W4847501, exp. 6/20/2027; and W4851589, exp. 7/8/2027.
- FDA recall number
- Z-0582-2025
- Quantity
- 46 devices
Distribution (FDA): Distribution was OUS only to Argentina, Brazil, and Israel.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Wilson-Cook Medical Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.