Wilson-Cook Medical Inc. recalls
5 notices naming Wilson-Cook Medical Inc. as the recalling firm, since December 11, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Wilson-Cook Medical Inc.: Brand Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT Product Name: HEMOSPRAY ENDOSCOPIC HEMOSTAT…
Hazard (FDA): Due to issue with endoscope adherence and inability to spray powder. Firm will provide an updated Instruction for Use (IFU).
FDAMedical devices
Wilson-Cook Medical Inc.: Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only
Hazard (FDA): Due to increased in complaints their is the potential for endoscopic clipping device to malfunction.
FDAMedical devices
Wilson-Cook Medical Inc.: Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code…
Hazard (FDA): Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
FDAMedical devices
Wilson-Cook Medical Inc.: Product Name: Nasal Jejunal Feeding Tube REF: NJFT-10
Hazard (FDA): Nasal feeding tube packaged without the nasal transfer tube component.
FDAMedical devices
Wilson-Cook Medical Inc.: Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat
Hazard (FDA): Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.
FDAMedical devices