What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
19 more codes are in the record below.
Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile
- Lot or code
- UDI-DI (GTIN): 05415067045768, Serial Numbers: 20078589, 20357219, 20357226, 20357685, 20358319, 20358343, 20358351, 20358363, 20358368, & 20358370.
- FDA recall number
- Z-0491-2025
- Quantity
- 1 US, 9 OUS
Abbott Navitor Transcatheter Aortic Heart Valve, 29 mm, NVTR-29, Sterile
- Lot or code
- UDI-DI (GTIN): 5415067031600, Serial Number: 20131478
- FDA recall number
- Z-0492-2025
- Quantity
- 1 OUS
Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile
- Lot or code
- UDI-DI (GTIN): 5415067045775, Serial Numbers: 20150441, 20156804, & 20170535
- FDA recall number
- Z-0493-2025
- Quantity
- 3 OUS
Abbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, Sterile
- Lot or code
- UDI-DI (GTIN): 5415067045799, Serial Numbers: 20158135, 20172203, 20183968, 20184552, 20184670, 20184675, 20190249, & 20213055.
- FDA recall number
- Z-0494-2025
- Quantity
- 8 OUS
Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
- Lot or code
- UDI-DI (GTIN): 5415067031594, Serial Numbers: 20110016, 20155472, 20155476, 20166304, & 20168050.
- FDA recall number
- Z-0495-2025
- Quantity
- 5 OUS
Distribution (FDA): New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Abbott Medical
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.