Abbott Medical: Merlin" Patient Care System
Hazard (FDA): Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
FDAMedical devices
Abbott Medical: Eterna SCS IPG and 1 more
Hazard (FDA): There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse Generators (IPGs), Model 62400…
FDAMedical devices
Abbott Medical: Aveir DR Leadless Pacemaker
Hazard (FDA): Due to a manufacturing issue, leadless pacemakers may have been exposed to higher than normal temperatures which may result in premature battery failure.
FDAMedical devices
Abbott Medical: The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a…
Hazard (FDA): Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the…
FDAMedical devices
Abbott Medical: Merlin@home Transmitter REF EX1100 The Merlin@home" and 1 more
Hazard (FDA): Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.
FDAMedical devices
Abbott Medical: Abbott Navitor Titan and 4 more
Hazard (FDA): A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or…