What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MICRO-X Rover Mobile X-ray System, # MXU-RV35
- Lot or code
- UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿
- FDA recall number
- Z-0664-2025
- Quantity
- 5
Distribution (FDA): Worldwide and US Nationwide Distribution: NJ, PR, CA
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Micro-X Ltd.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.