Micro-X Ltd. recalls
7 notices naming Micro-X Ltd. as the recalling firm, since December 4, 2024. A notice covers specific products and lots, not every product of the firm.
Hazard and remedy text is quoted from the agency, unedited. The tag names the kind of risk with a word from that text. Every notice links to its official page.
Micro-X Ltd.: Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71
Hazard (FDA): X-Ray generator may malfunction resulting in the X-Ray being inoperable.
FDAMedical devices
Micro-X Ltd.: Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 and 1 more
Hazard (FDA): Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
FDAMedical devices
Micro-X Ltd.: Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71… and 1 more
Hazard (FDA): Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
FDAMedical devices
Micro-X Ltd.: MICRO-X Rover Mobile X-ray System
Hazard (FDA): Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
FDAMedical devices
Micro-X Ltd.: MICRO-X Rover Mobile X-ray System
Hazard (FDA): Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
FDAMedical devices
Micro-X Ltd.: MICRO-X Rover Mobile X-ray System
Hazard (FDA): During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a…
FDAMedical devices
Micro-X Ltd.: MICRO-X Rover Mobile X-ray System
Hazard (FDA): Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to…
FDAMedical devices