Medline Industries, LP Issues Nationwide Recall of Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits

FDA recall n-medline-industries-lp-issues-nationwide-recall-hudson-rci-neonatalinfant-heated- ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Note from Recalls Watch, not agency text

Why it was recalled

Due to heated wire connector (pigtail) at the Neptune humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage
Text from FDA, unedited.

The recalled product

On July 13, 2026, Medline Industries, LP initiated a nationwide recall of Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits. Medline received reports of sporadic failures involving the electrical connector of certain neonatal and infant heated wire breathing circuits. In affected circuits, t
Text from FDA, unedited.

Company announcement, as FDA publishes it

On July 13, 2026, Medline Industries, LP initiated a nationwide recall of Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits. Medline received reports of sporadic failures involving the electrical connector of certain neonatal and infant heated wire breathing circuits. In affected circuits, the heated wire connector (pigtail) at the Neptune humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. To date, reported events have occurred infrequently relative to the number of devices distributed, with the observed occurrence rate of 0.011% across the distributed population. Refer to the below photographs of the product and a defect that was reported to Medline.

The Neptune humidifier is designed with safety features that continually monitor the heated wire circuit and automatically discontinues power to the heated wires when electrical abnormalities are detected. When power to the heated wires is disconnected, respiratory gas flow to the patient continues. The disconnection of power is limited to the heated wire function.

Continued delivery of inadequately heated or unconditioned gas can desiccate and damage the airway epithelium, predisposing the infant to bronchospasm, inspissated secretions, and airway obstruction. Potential continued loss of heated wire function may result in increased circuit condensation (rainout), which can cause inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the infant's airway, reduced humidity and delivery of cold, dry gas increases the risk of cold stress, leading to elevated oxygen consumption and worsening respiratory distress.

Consumers who have the affected Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits should stop using and discard the device. Consumers should not return product to Medline.

Medline voluntarily recalled product after thorough investigation of reports received for this issue and notified the FDA of this action. At this time, Medline is not aware of any serious health consequences which have occurred due to use of the device.

Recalled Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits were manufactured on or before July 8, 2026 and distributed from June 28, 2021, to July 31, 2026.

Hudson RCI Neonatal/Infant Heated Wire Breathing Circuits within the scope of the recall can be identified by their product labeling, which includes the name of the device, item number, and manufacturing date. The labeling of products affected by the recall include either a Teleflex or Medline item number, depending on the manufacturing date. Affected products manufactured before January 2024 display a Teleflex label and item number, while products manufactured starting January 2024 or later display a Medline label and item number. Please refer to the example labels below to identify the manufacturing date (red) and item number (blue) on the product label.

The following item numbers are included in the scope of the recall:

All product manufactured on or after July 9, 2026 are not under scope of the recall and may continue to be used.

Medline is notifying its distributors and customers by email and/or first-class mail and is arranging for credit of all recalled product(s).

Consumers with questions regarding this recall can contact Medline by phone at 866-359-1704 Monday through Friday between the hours of 8am and 5pm CST or recalls@medline.com . Consumers should contact their physician or healthcare provider if they have experienced any problems that may be related to using this medical device.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Complete and submit the report Online : www.fda.gov/medwatch/report.htm

Regular Mail or Fax : Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-3321088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178

Text from FDA (the company’s words), unedited.
  • HUD78001KIT; 780-01KIT

    Lot or code
    Medline Item Number: HUD78001KIT; Teleflex Item Number: 780-01KIT
  • HUD78007; 780-07

    Lot or code
    Medline Item Number: HUD78007; Teleflex Item Number: 780-07
  • HUD78007KIT; 780-07KIT

    Lot or code
    Medline Item Number: HUD78007KIT; Teleflex Item Number: 780-07KIT
  • HUD78008KIT; 780-08KIT

    Lot or code
    Medline Item Number: HUD78008KIT; Teleflex Item Number: 780-08KIT
  • HUD78009; 780-09

    Lot or code
    Medline Item Number: HUD78009; Teleflex Item Number: 780-09
  • HUD78009KIT; 780-09KIT

    Lot or code
    Medline Item Number: HUD78009KIT; Teleflex Item Number: 780-09KIT
  • HUD78013; 780-13

    Lot or code
    Medline Item Number: HUD78013; Teleflex Item Number: 780-13
  • HUD78014; 780-14

    Lot or code
    Medline Item Number: HUD78014; Teleflex Item Number: 780-14
  • HUD78015; 780-15

    Lot or code
    Medline Item Number: HUD78015; Teleflex Item Number: 780-15
  • HUD78017KIT; 780-17KIT

    Lot or code
    Medline Item Number: HUD78017KIT; Teleflex Item Number: 780-17KIT
  • HUD78018; 780-18

    Lot or code
    Medline Item Number: HUD78018; Teleflex Item Number: 780-18
  • HUD78018KIT; 780-18KIT

    Lot or code
    Medline Item Number: HUD78018KIT; Teleflex Item Number: 780-18KIT
  • HUD78019; 780-19

    Lot or code
    Medline Item Number: HUD78019; Teleflex Item Number: 780-19
  • HUD88001KIT; 880-01KIT

    Lot or code
    Medline Item Number: HUD88001KIT; Teleflex Item Number: 880-01KIT
  • HUD88013KIT; 880-13KIT

    Lot or code
    Medline Item Number: HUD88013KIT; Teleflex Item Number: 880-13KIT
  • HUD88014KIT; 880-14KIT

    Lot or code
    Medline Item Number: HUD88014KIT; Teleflex Item Number: 880-14KIT
  • HUD88015KIT; 880-15KIT

    Lot or code
    Medline Item Number: HUD88015KIT; Teleflex Item Number: 880-15KIT
  • HUD88017KIT; 880-17KIT

    Lot or code
    Medline Item Number: HUD88017KIT; Teleflex Item Number: 880-17KIT

Source

Issued by
U.S. Food and Drug Administration
Firm
Medline Industries, LP (other notices of this firm)

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Medline Industries, LP, Month: August 2026, Week 2026-W32. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Due to heated wire connector (pigtail) at the Neptune humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published this company announcement on August 5, 2026.

Other notices of Medline Industries, LP