Skip to content
FDAMedical devices

BD Provides Update on Voluntary Recall of Certain BD Alaris™ Pump Infusion Sets

FDA recall n-bd-provides-update-voluntary-recall-certain-bd-alaristm-pump-infusion-sets ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

What the agency says to do

FDA’s record gives no instruction for consumers.

Note from Recalls Watch, not agency text

Why it was recalled

Pump performance variations compared to the performance described in the user manual could impact infusion delivery
Text from FDA, unedited.

The recalled product

Codes in the record. Click one to copy it.

FRANKLIN LAKES, N.J. (Sept. 12, 2025) – BD (Becton, Dickinson and Company) (NYSE: BDX), a leading global medical technology company, today expanded their Class I voluntary recall initiated in the United States on July 8, 2025 and expanded on July 17, 2025 to inform customers of worst-case performanc
Text from FDA, unedited.

Company announcement, as FDA publishes it

FRANKLIN LAKES, N.J. (Sept. 12, 2025) – BD (Becton, Dickinson and Company) (NYSE: BDX), a leading global medical technology company, today expanded their Class I voluntary recall initiated in the United States on July 8, 2025 and expanded on July 17, 2025 to inform customers of worst-case performance, under certain use cases, for the BD Alaris™ Pump Module model 8100 when used with a subset of compatible pump infusion sets, as compared to the performance described in the user manual. The update includes an expansion of 15 additional impacted pump infusion sets included in the recall, as well as new corrective actions and risk information. The 15 additional sets had been previously discontinued but may remain in inventory based on the expiration date.

This issue was identified through internal testing and, to date, BD has not received any complaints associated with this issue. The affected products have the potential to result in patient death or serious adverse events, with a particular higher risk to vulnerable patient populations such as neonates and critically ill patients. The recall has been designated as a Class I recall by the U.S. Food and Drug Administration (FDA).

Pump performance variations as compared to the performance described in the user manual could impact infusion delivery in the following ways:

Flow rate and loading bolus dose accuracy (over/under infusion),

Upstream and downstream occlusion alarm delay,

and post-occlusion bolus volume (POBV) over infusion,

which could adversely impact infusion performance and dose accuracy. The severity and nature of these outcomes depend on the type of medication, fluid, or infusate being administered, as well as the individual patient’s condition.

The deviations in performance from the previously published ranges are attributable to these pump infusion sets' design features, such as filters and other in-line components. The products included in the recall were distributed nationwide in the United States, including Guam and Puerto Rico, and in Canada, Belgium and South Africa.

In an update sent to all customers on September 11, 2025, BD shared updated corrective actions, including risk mitigations and recommendations for alternate BD Alaris™ Pump infusion sets to use when available and clinically appropriate. The customer update also included a list of the 15 additional impacted pump infusion sets included in the recall. Information about this recall is available on BD's website or by calling BD at 1-888-562-6018.

Customers should report any complaints experienced with the use of this product to the BD Complaint Center at 1-844-823-5433, Monday – Friday between the hours of 9 a.m. and 6 p.m. ET or by emailing productcomplaints@bd.com .

FDA MedWatch Reporting

Adverse reactions/events experienced with the use of this product should be reported to the FDA's MedWatch Program by:

Online: Complete and submit the report at www.fda.gov/medwatch/report.htm

Call 1-800-FDA-1088 (1-800-332-1088)

Regular Mail: Download form www.fda.gov/MedWatch/getforms.htm or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.

About BD

BD is one of the largest global medical technology companies in the world and is advancing the world of health by improving medical discovery, diagnostics and the delivery of care. The company supports the heroes on the frontlines of health care by developing innovative technology, services and solutions that help advance both clinical therapy for patients and clinical process for health care providers. BD and its more than 70,000 employees have a passion and commitment to help enhance the safety and efficiency of clinicians' care delivery process, enable laboratory scientists to accurately detect disease and advance researchers' capabilities to develop the next generation of diagnostics and therapeutics. BD has a presence in virtually every country and partners with organizations around the world to address some of the most challenging global health issues. By working in close collaboration with customers, BD can help enhance outcomes, lower costs, increase efficiencies, improve safety and expand access to health care. For more information on BD, please visit bd.com or connect with us on LinkedIn at www.linkedin.com/company/bd1/ , X (formerly Twitter) @BDandCo or Instagram @becton_dickinson .

Contacts:

Media: Fallon McLoughlin Director, Public Relations 201.258.0361 fallon.mcloughlin@bd.com

Investors: Adam Reiffe Sr. Director, Investor Relations 201.847.6927 adam.reiffe@bd.com

Text from FDA (the company’s words), unedited.
  • BD Alaris BD Alaris Pump Module model 8100 used with a subset of compatible pump infusion sets

Source

Issued by
U.S. Food and Drug Administration
Firm
BD

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: BD, Month: September 2025, Week 2025-W37. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Pump performance variations compared to the performance described in the user manual could impact infusion delivery”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

Which products are included?

FDA states: “FRANKLIN LAKES, N.J. (Sept. 12, 2025) – BD (Becton, Dickinson and Company) (NYSE: BDX), a leading global medical technology company, today expanded their Class I voluntary recall initiated in the United States on July 8, 2025 and expanded on July 17, 2025 to inform customers of worst-case performanc”

When was this recall published?

FDA published this company announcement on September 12, 2025.