Why it was recalled
Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
14 more codes are in the record below.
FORWARD Gunshot Wound Kit; Model: AFT-GSW;
- Lot or code
- Model: ATF-GSW; UDI-DI: 00850056748063; Lot Numbers: IMC-00000024, IMC-00000034, IMC-00000053, IMC-00000056, IMC-00000076, IMC-00000081, IMC-00000082, IMC-00000083, IMC-00000084, IMC-00000085, IMC-00000094, IMC-00000148, IMC-00000149, IMC-00000152, IMC-00000153, IMC-00000154, IMC-00000173, IMC-00000174, IMC-00000179, IMC-00000186, IMC-00000187;
- FDA recall number
- Z-0147-2027
- Quantity
- 1033
Distribution (FDA): US Nationwide distribution in the states of GA, VA.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- INTEGRATED MEDCRAFT LLC
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.