FDAMedical devicesClass IIOngoing

FDA recall 99941: NATUS MEDICAL INCORPORATED: Embletta MPR PG Multi Parameter Recorder Sleep Monitor

FDA recall 99941 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

247 more codes are in the record below.

  • Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011

    Lot or code
    UDI-DI: 00382830023764; Serial Numbers: EBMPG-C1013678 EBMPG-C1013679 EBMPG-C1013682 EBMPG-C1013690 EBMPG-C1013691 EBMPG-C1013697 EBMPG-C1013700 EBMPG-C1013701 EBMPG-C1013703 EBMPG-C1013704 EBMPG-C1013707 EBMPG-C1013708 EBMPG-C1013748 EBMPG-C1013778 EBMPG-C1013779 EBMPG-C1013876 EBMPG-C1013566 EBMPG-C1013620 EBMPG-C1013632 EBMPG-C1013594 EBMPG-C1013596 EBMPG-C1013597 EBMPG-C1013599 EBMPG-C1013602 EBMPG-C1013614 EBMPG-C1013615 EBMPG-C1013617 EBMPG-C1013618 EBMPG-C1013414 EBMPG-C1013415 EBMPG-C1013416 EBMPG-C1013417 EBMPG-C1013448 EBMPG-C1013418 EBMPG-C1013591 EBMPG-C1013621 EBMPG-C1013629 EBMPG-C1013634 EBMPG-C1013636 EBMPG-C1013637 EBMPG-C1013643 EBMPG-C1013648 EBMPG-C1013655 EBMPG-C1013665 EBMPG-C1013609 EBMPG-C1013650 EBMPG-C1013652 EBMPG-C1013666 EBMPG-C1013670 EBMPG-C1013649 EBMPG-C1013293 EBMPG-C1013376 EBMPG-C1013387 EBMPG-C1013392 EBMPG-C1013398 EBMPG-C1013616 EBMPG-C1013633 EBMPG-C1013630 EBMPG-C1013638 EBMPG-C1013627 EBMPG-C1013541 EBMPG-C1013661 EBMPG-C1013667 EBMPG-C1013495 EBMPG-C1013572 EBMPG-C1013577 EBMPG-C1013578 EBMPG-C1013595 EBMPG-C1013610 EBMPG-C1013642 EBMPG-C1013653 EBMPG-C1013723 EBMPG-C1013734 EBMPG-C1013587 EBMPG-C1013590 EBMPG-C1013606 EBMPG-C1013611 EBMPG-C1013613 EBMPG-C1013641 EBMPG-C1013676 EBMPG-C1013542 EBMPG-C1013574 EBMPG-C1013583 EBMPG-C1013584 EBMPG-C1013672 EBMPG-C1013673 EBMPG-C1013680 EBMPG-C1013681 EBMPG-C1013683 EBMPG-C1013685 EBMPG-C1013686 EBMPG-C1013687 EBMPG-C1013688 EBMPG-C1013689 EBMPG-C1013693 EBMPG-C1013694 EBMPG-C1013695…
    FDA recall number
    Z-0319-2027
    Quantity
    288 units

    The text above is the beginning of the record; 2,340 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide distribution - US Nationwide and the countries of Canada, China, Italy, Portugal.

Source

Issued by
U.S. Food and Drug Administration
Firm
NATUS MEDICAL INCORPORATED
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: November 2026, Week 2026-W46. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99941 on November 11, 2026.