What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
26 more codes are in the record below.
EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
- Lot or code
- UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351
- FDA recall number
- Z-0301-2027
- Quantity
- 33
Distribution (FDA): Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- BioMerieux SA
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.