Skip to content
FDAMedical devicesClass IOngoing

FDA recall 99851: Tandem Diabetes Care, Inc.: Tandem Mobi Insulin Pump

FDA recall 99851 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
Text from FDA, unedited.

FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

13 more codes are in the record below.

  • Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811, 1014813, 1014936, 1016911, 1016912, 1016913, 1016914, 1018075, 1018077, 1018078

    Lot or code
    UDI-DI: 00389152075013 00389152271910 00389152350110 00389152365510 00389152365619 00389152370019 00389152480114 00389152481111 00389152481319. All affected pump models and serial numbers may be affected. Pump serial numbers with frequent false-positive occlusion alarms to be exchanged: 1198599 1199144 1201048 1201777 1202107 1202156 1204500 1204935 1207112 1209405 1209862 1211585 1218143 1220121 1220201 1220470 1220471 1221343 1221347 1221775 1221783 1222111 1223346 1223487 1223488 1223886 1223906 1223940 1224363 1224978 1225017 1225024 1225158 1225171 1225194 1225204 1225589 1225602 1225682 1226359 1226360 1226365 1226671 1226721 1226748 1226859 1226989 1227008 1227021 1227227 1227375 1227442 1229300 1229807 1229969 1230064 1231839 1233326 1233515 1233517 1233562 1233617 1233665 1233701 1233711 1234069 1234080 1234136 1234188 1234240 1234244 1234521 1234597 1234601 1234703 1234715 1234717 1234718 1234732 1235069 1235124 1235131 1235132 1235138 1235193 1235195 1235204 1235335 1236661 1236815 1236825 1236879 1236945 1237266 1237283 1237337 1237722 1237727 1238064 1238085 1238466 1238517 1238533 1238538 1238742 1238801 1238806 1238820 1238851 1238882 1238899 1238923 1239003 1239011 1239016 1239048 1239054 1239057 1239058 1239094 1239104 1239120 1239489 1239496 1239518 1239527 1239560 1239756 1239762 1239794 1239812 1239814 1239815 1239845 1239877 1239884 1239885 1240021 1241967 1242353 1242365 1242366 1242374 1242450 1242747 1243800 1243835 1243889 1243922 1244195 1244451…
    FDA recall number
    Z-0316-2027
    Quantity
    80,938 (77,559 US, 3,379 OUS)

    The text above is the beginning of the record; 35,632 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of CA, FL, NC, OK, UT, WA, IL, PA, MN, VA, NY, IN, HI, OR, GA, MI, KS, CO, WI, SC, null, TX, NE, AZ, OH, NM, NV, MA, SD, IA, TN, ID, MD, RI, CT, NJ, ME, KY, MT, WV, VT, MS, AR, AL, DC, LA, MO, NH, PR, ND, AK, DE, WY and the country of Canada.

Source

Issued by
U.S. Food and Drug Administration
Firm
Tandem Diabetes Care, Inc. (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Tandem Diabetes Care, Inc., Month: November 2026, Week 2026-W46. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm…”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99851 on November 11, 2026.

Other notices of Tandem Diabetes Care, Inc.