Why it was recalled
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
- Lot or code
- Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;
- FDA recall number
- Z-0117-2027
- Quantity
- 54
Distribution (FDA): US Nationwide distribution in the states of CA, FL, GA, KY, LA, NJ, OH, PA, SC.