What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
3 more codes are in the record below.
Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249
- Lot or code
- REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN
- FDA recall number
- Z-0119-2027
- Quantity
- 8
Distribution (FDA): Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Intellijoint Surgical, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.