What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
- Lot or code
- REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
- FDA recall number
- Z-0159-2027
- Quantity
- 1366
Distribution (FDA): Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Microgenics Corporation
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.