Why it was recalled
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
158 more codes are in the record below.
Computed Tomography X-ray System, Model: uCT 530 with software
- Lot or code
- UDI-DI: 06971576831012. Serial Numbers/Software: 300174/R001.19.5.1839204-Re3300SP1H1_P4, 304006/R001.17.6.1607336-Re3100SP4, 304007/R001.19.4.1810621-Re3300SP1H1_P3, 304008/R001.19.4.1810621-Re3300SP1H1_P3
- FDA recall number
- Z-3277-2026
- Quantity
- 4
Computed Tomography X-ray System, Model: uCT 550 with software
- Lot or code
- UDI-DI: 06971576831036. Serial Numbers/Software: 353021/R001.19.6.1857994-Re3300SP1H1_P5, 353045/R001.19.6.1857994-Re3300SP1H1_P5, 353049/R001.19.6.1857994-Re3300SP1H1_P5, 354001/R001.19.6.1857994-Re3300SP1H1_P5, 354004/R001.19.6.1857994-Re3300SP1H1_P5, 354005/R001.19.6.1857994-Re3300SP1H1_P5, 354007/R001.19.6.1857994-Re3300SP1H1_P5, 354009/R001.19.6.1857994-Re3300SP1H1_P5, 354013/R001.19.6.1857994-Re3300SP1H1_P5, 354015/R001.19.6.1857994-Re3300SP1H1_P5, 354016/R001.19.6.1857994-Re3300SP1H1_P5, 354018/R001.17.6.1607336-Re3100SP4, 354020/R001.17.6.1607336-Re3100SP4, 354021/R001.17.6.1607336-Re3100SP4, 354022/R001.19.4.1810621-Re3300SP1H1_P3, 354023/R001.19.6.1857994-Re3300SP1H1_P5, 354024/R001.19.6.1857994-Re3300SP1H1_P5, 354025/R001.19.6.1857994-Re3300SP1H1_P5, 354026/R001.19.6.1857994-Re3300SP1H1_P5, 354027/R001.19.6.1857994-Re3300SP1H1_P5, 354028/R001.19.6.1857994-Re3300SP1H1_P5, 354029/R001.19.6.1857994-Re3300SP1H1_P5, 354030/R001.19.6.1857994-Re3300SP1H1_P5
- FDA recall number
- Z-3278-2026
- Quantity
- 23
Computed Tomography X-ray System, Model: uCT 760 with software
- Lot or code
- UDI-DI: 06971576831043. Serial Number/Software: 600271/R003.8.4.1810621-Re3300SP1H1_P3
- FDA recall number
- Z-3279-2026
- Quantity
- 1
Computed Tomography X-ray System, Model: uCT 780 with software
- Lot or code
- UDI-DI: 06971576831050. Serial Number/Software: 684001/R001.17.6.1607336-Re3100SP4, 684002/R001.17.6.1607336-Re3100SP4, 684003/R003.8.4.1810621-Re3300SP1H1_P3, 684004/R003.8.4.1810621-Re3300SP1H1_P3, 684005/R003.8.4.1810621-Re3300SP1H1_P3, 684006/R003.8.4.1810621-Re3300SP1H1_P3, 684007/R003.8.4.1810621-Re3300SP1H1_P3, 684008/R003.8.4.1810621-Re3300SP1H1_P3, 684009/R003.8.4.1810621-Re3300SP1H1_P3, 684010/R003.8.4.1810621-Re3300SP1H1_P3, 684011/R003.8.4.1810621-Re3300SP1H1_P3, 684012/R003.9.0.1776134-Re-3300SP2, 684013/R003.11.0.1897081-Re3310SP1, 684014/R003.11.0.1897081-Re3310SP1
- FDA recall number
- Z-3280-2026
- Quantity
- 14
Computed Tomography X-ray System, Model: uCT ATLAS with software
- Lot or code
- UDI-DI: 06971576831104. Serial Number/Software: 863001/R001.17.6.1607336-Re3100SP4, 863002/R001.17.6.1607336-Re3100SP4, 863004/R001.17.6.1607336-Re3100SP4, 863005/R001.17.6.1607336-Re3100SP4, 863007/R001.21.0.1897081-Re3310SP1, 863008/R001.21.0.1897081-Re3310SP1, 869001/R001.22.0.2012850-Re3320, 869002/R001.19.4.1810621-Re3300SP1H1_P3, 869003/R001.21.0.1897081-Re3310SP1, 869004/R001.21.0.1897081-Re3310SP1, 11CT863010/R001.17.6.1607336-Re3100SP4
- FDA recall number
- Z-3281-2026
- Quantity
- 11
Computed Tomography X-ray System, Model: uCT ATLAS Astound with software
- Lot or code
- UDI-DI: 06971576831340. Serial Number/Software: 803011/R002.9.0.1897081-Re3310SP1, 803012/R002.4.6.1607336-Re3100SP4, 803013/R002.4.6.1607336-Re3100SP4, 803014/R002.9.0.1897081-Re3310SP1, 803016/R002.4.6.1607336-Re3100SP4, 803017/R002.9.0.1897081-Re3310SP1, 803018/R002.9.0.1897081-Re3310SP1, 803019/R002.4.6.1607336-Re3100SP4, 803020/R002.9.0.1897081-Re3310SP1, 803022/R002.9.0.1897081-Re3310SP1, 803023/R002.9.0.1897081-Re3310SP1, 803024/R002.9.0.1897081-Re3310SP1, 809001/R002.8.1.1840891-Re3310A_P1, 809002/R002.9.0.1897081-Re3310SP1, 809003/R002.8.1.1840891-Re3310A_P1, 809004/R002.4.6.1607336-Re3100SP4, 809005/R002.4.6.1607336-Re3100SP4, 809006/R002.8.1.1840891-Re3310A_P1, 809007/R002.4.7.1671762-Re3100SP4_P1, 809008/R002.4.7.1671762-Re3100SP4_P1, 809009/R002.4.7.1671762-Re3100SP4_P1, 809010/R002.4.7.1671762-Re3100SP4_P1, 809011/R002.4.7.1671762-Re3100SP4_P1, 809012/R002.4.7.1671762-Re3100SP4_P1, 809013/R002.4.7.1671762-Re3100SP4_P1, 809014/R002.7.4.1810621-Re3300SP1H1_P3, 809015/R002.7.4.1810621-Re3300SP1H1_P3, 809016/R002.9.0.1897081-Re3310SP1, 809017/R002.9.0.1897081-Re3310SP1, 809018/R002.7.4.1810621-Re3300SP1H1_P3, 809019/R002.7.4.1810621-Re3300SP1H1_P3, 809020/R002.9.0.1897081-Re3310SP1, 809021/R002.7.4.1810621-Re3300SP1H1_P3, 809022/R002.7.4.1810621-Re3300SP1H1_P3, 809023/R002.7.4.1810621-Re3300SP1H1_P3, 809024/R002.7.4.1810621-Re3300SP1H1_P3, 809025/R002.7.4.1810621-Re3300SP1H1_P3, 809026/R002.7.4.1810621-Re3300SP1H1_P3, 809027/R002.7.4.1810621-Re3300SP1H1_P3,…
- FDA recall number
- Z-3282-2026
- Quantity
- 53
The text above is the beginning of the record; 509 more characters are in the official notice. See the full list in the official notice (opens in a new tab)
Positron Emission Tomography and Computed Tomography System, Model: uMI 550 with software
- Lot or code
- UDI-DI: 06971576832026. Serial Number/Software: 200045/R001.19.0.1812568, 200163/R001.19.3.2011750, 200164/R001.18.0.1683822, 200165/R001.18.0.1683822, 200166/R001.18.0.1683822, 200167/R001.18.0.1683822, 202009/R001.19.3.2011750, 202016/R001.19.0.1812568, 202018/R001.19.0.1812568, 202019/R001.19.0.1812568, 202020/R001.19.1.1929746, 202021/R001.19.1.1929746, 202022/R001.19.1.1929746, 230040/R001.19.3.2011750, 230048/R001.19.0.1812568, 230106/R001.19.0.1812568, 232038/R001.19.0.1812568, 232039/R001.19.0.1812568, 232040/R001.19.0.1812568, 232041/R001.19.0.1812568, 232042/R001.19.0.1812568, 232043/R001.19.0.1812568, 232044/R001.19.0.1812568, 232045/R001.19.0.1812568, 232046/R001.19.0.1812568, 232047/R001.19.0.1812568, 232048/R001.19.1.1929746, 232049/R001.19.1.1929746, 232050/R001.19.1.1929746, 232051/R001.19.1.1929746, 202023/R001.19.1.1929746, 202024/R001.19.1.1929746, 232052/R001.19.1.1929746, 232053/R001.19.1.1929746
- FDA recall number
- Z-3283-2026
- Quantity
- 34
Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama 35 with software
- Lot or code
- UDI-DI: 06971576832170. Serial Number: 11MI410003/R001.19.4.1909524, 11MI410004/R001.19.4.1909524, 11MI410005/R001.19.4.1909524, 20MI410001/R001.19.4.1909524, 20MI410002/R001.19.4.1909524, 20MI410003/R001.21.0.1993334, 20MI410004/R001.19.4.1909524
- FDA recall number
- Z-3284-2026
- Quantity
- 7
Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software
- Lot or code
- UDI-DI: 06971576832187. Serial Number: 11MIP00007/R001.17.5.1785268, 11MIP00009/R001.18.2.1967222, 11MIP00010/R001.18.2.1967222
- FDA recall number
- Z-3285-2026
- Quantity
- 3
Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software
- Lot or code
- UDI-DI: 06971576832231. Serial Number/Software: 11MI500004/R002.4.0.1872339, 11MI500005/R002.4.0.1872339, 11MI500006/R002.4.0.1872339
- FDA recall number
- Z-3286-2026
- Quantity
- 3
Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo S with software
- Lot or code
- UDI-DI: 06971576832255. Serial Number/Software: 11MI510005/R003.2.0.1872339, 20MI510001/R003.2.0.1872339, 20MI510002/R003.2.0.1872339, 20MI510003/R003.2.0.1872339, 20MI510004/R003.2.0.1872339, 20MI510005/R003.2.0.1872339
- FDA recall number
- Z-3287-2026
- Quantity
- 6
Positron Emission Tomography and Computed Tomography System, Model: uMI 780
- Lot or code
- UDI-DI: 06971576832033. Serial Number/Software: 300098/R001. 19.0.1812568
- FDA recall number
- Z-3288-2026
- Quantity
- 1
Distribution (FDA): US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Shanghai United Imaging Healthcare Co., Ltd.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.