Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99731: Shanghai United Imaging Healthcare Co., Ltd.: Computed Tomography X-ray System and 11 more

FDA recall 99731 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

158 more codes are in the record below.

  • Computed Tomography X-ray System, Model: uCT 530 with software

    Lot or code
    UDI-DI: 06971576831012. Serial Numbers/Software: 300174/R001.19.5.1839204-Re3300SP1H1_P4, 304006/R001.17.6.1607336-Re3100SP4, 304007/R001.19.4.1810621-Re3300SP1H1_P3, 304008/R001.19.4.1810621-Re3300SP1H1_P3
    FDA recall number
    Z-3277-2026
    Quantity
    4
  • Computed Tomography X-ray System, Model: uCT 550 with software

    Lot or code
    UDI-DI: 06971576831036. Serial Numbers/Software: 353021/R001.19.6.1857994-Re3300SP1H1_P5, 353045/R001.19.6.1857994-Re3300SP1H1_P5, 353049/R001.19.6.1857994-Re3300SP1H1_P5, 354001/R001.19.6.1857994-Re3300SP1H1_P5, 354004/R001.19.6.1857994-Re3300SP1H1_P5, 354005/R001.19.6.1857994-Re3300SP1H1_P5, 354007/R001.19.6.1857994-Re3300SP1H1_P5, 354009/R001.19.6.1857994-Re3300SP1H1_P5, 354013/R001.19.6.1857994-Re3300SP1H1_P5, 354015/R001.19.6.1857994-Re3300SP1H1_P5, 354016/R001.19.6.1857994-Re3300SP1H1_P5, 354018/R001.17.6.1607336-Re3100SP4, 354020/R001.17.6.1607336-Re3100SP4, 354021/R001.17.6.1607336-Re3100SP4, 354022/R001.19.4.1810621-Re3300SP1H1_P3, 354023/R001.19.6.1857994-Re3300SP1H1_P5, 354024/R001.19.6.1857994-Re3300SP1H1_P5, 354025/R001.19.6.1857994-Re3300SP1H1_P5, 354026/R001.19.6.1857994-Re3300SP1H1_P5, 354027/R001.19.6.1857994-Re3300SP1H1_P5, 354028/R001.19.6.1857994-Re3300SP1H1_P5, 354029/R001.19.6.1857994-Re3300SP1H1_P5, 354030/R001.19.6.1857994-Re3300SP1H1_P5
    FDA recall number
    Z-3278-2026
    Quantity
    23
  • Computed Tomography X-ray System, Model: uCT 760 with software

    Lot or code
    UDI-DI: 06971576831043. Serial Number/Software: 600271/R003.8.4.1810621-Re3300SP1H1_P3
    FDA recall number
    Z-3279-2026
    Quantity
    1
  • Computed Tomography X-ray System, Model: uCT 780 with software

    Lot or code
    UDI-DI: 06971576831050. Serial Number/Software: 684001/R001.17.6.1607336-Re3100SP4, 684002/R001.17.6.1607336-Re3100SP4, 684003/R003.8.4.1810621-Re3300SP1H1_P3, 684004/R003.8.4.1810621-Re3300SP1H1_P3, 684005/R003.8.4.1810621-Re3300SP1H1_P3, 684006/R003.8.4.1810621-Re3300SP1H1_P3, 684007/R003.8.4.1810621-Re3300SP1H1_P3, 684008/R003.8.4.1810621-Re3300SP1H1_P3, 684009/R003.8.4.1810621-Re3300SP1H1_P3, 684010/R003.8.4.1810621-Re3300SP1H1_P3, 684011/R003.8.4.1810621-Re3300SP1H1_P3, 684012/R003.9.0.1776134-Re-3300SP2, 684013/R003.11.0.1897081-Re3310SP1, 684014/R003.11.0.1897081-Re3310SP1
    FDA recall number
    Z-3280-2026
    Quantity
    14
  • Computed Tomography X-ray System, Model: uCT ATLAS with software

    Lot or code
    UDI-DI: 06971576831104. Serial Number/Software: 863001/R001.17.6.1607336-Re3100SP4, 863002/R001.17.6.1607336-Re3100SP4, 863004/R001.17.6.1607336-Re3100SP4, 863005/R001.17.6.1607336-Re3100SP4, 863007/R001.21.0.1897081-Re3310SP1, 863008/R001.21.0.1897081-Re3310SP1, 869001/R001.22.0.2012850-Re3320, 869002/R001.19.4.1810621-Re3300SP1H1_P3, 869003/R001.21.0.1897081-Re3310SP1, 869004/R001.21.0.1897081-Re3310SP1, 11CT863010/R001.17.6.1607336-Re3100SP4
    FDA recall number
    Z-3281-2026
    Quantity
    11
  • Computed Tomography X-ray System, Model: uCT ATLAS Astound with software

    Lot or code
    UDI-DI: 06971576831340. Serial Number/Software: 803011/R002.9.0.1897081-Re3310SP1, 803012/R002.4.6.1607336-Re3100SP4, 803013/R002.4.6.1607336-Re3100SP4, 803014/R002.9.0.1897081-Re3310SP1, 803016/R002.4.6.1607336-Re3100SP4, 803017/R002.9.0.1897081-Re3310SP1, 803018/R002.9.0.1897081-Re3310SP1, 803019/R002.4.6.1607336-Re3100SP4, 803020/R002.9.0.1897081-Re3310SP1, 803022/R002.9.0.1897081-Re3310SP1, 803023/R002.9.0.1897081-Re3310SP1, 803024/R002.9.0.1897081-Re3310SP1, 809001/R002.8.1.1840891-Re3310A_P1, 809002/R002.9.0.1897081-Re3310SP1, 809003/R002.8.1.1840891-Re3310A_P1, 809004/R002.4.6.1607336-Re3100SP4, 809005/R002.4.6.1607336-Re3100SP4, 809006/R002.8.1.1840891-Re3310A_P1, 809007/R002.4.7.1671762-Re3100SP4_P1, 809008/R002.4.7.1671762-Re3100SP4_P1, 809009/R002.4.7.1671762-Re3100SP4_P1, 809010/R002.4.7.1671762-Re3100SP4_P1, 809011/R002.4.7.1671762-Re3100SP4_P1, 809012/R002.4.7.1671762-Re3100SP4_P1, 809013/R002.4.7.1671762-Re3100SP4_P1, 809014/R002.7.4.1810621-Re3300SP1H1_P3, 809015/R002.7.4.1810621-Re3300SP1H1_P3, 809016/R002.9.0.1897081-Re3310SP1, 809017/R002.9.0.1897081-Re3310SP1, 809018/R002.7.4.1810621-Re3300SP1H1_P3, 809019/R002.7.4.1810621-Re3300SP1H1_P3, 809020/R002.9.0.1897081-Re3310SP1, 809021/R002.7.4.1810621-Re3300SP1H1_P3, 809022/R002.7.4.1810621-Re3300SP1H1_P3, 809023/R002.7.4.1810621-Re3300SP1H1_P3, 809024/R002.7.4.1810621-Re3300SP1H1_P3, 809025/R002.7.4.1810621-Re3300SP1H1_P3, 809026/R002.7.4.1810621-Re3300SP1H1_P3, 809027/R002.7.4.1810621-Re3300SP1H1_P3,…
    FDA recall number
    Z-3282-2026
    Quantity
    53

    The text above is the beginning of the record; 509 more characters are in the official notice. See the full list in the official notice (opens in a new tab)

  • Positron Emission Tomography and Computed Tomography System, Model: uMI 550 with software

    Lot or code
    UDI-DI: 06971576832026. Serial Number/Software: 200045/R001.19.0.1812568, 200163/R001.19.3.2011750, 200164/R001.18.0.1683822, 200165/R001.18.0.1683822, 200166/R001.18.0.1683822, 200167/R001.18.0.1683822, 202009/R001.19.3.2011750, 202016/R001.19.0.1812568, 202018/R001.19.0.1812568, 202019/R001.19.0.1812568, 202020/R001.19.1.1929746, 202021/R001.19.1.1929746, 202022/R001.19.1.1929746, 230040/R001.19.3.2011750, 230048/R001.19.0.1812568, 230106/R001.19.0.1812568, 232038/R001.19.0.1812568, 232039/R001.19.0.1812568, 232040/R001.19.0.1812568, 232041/R001.19.0.1812568, 232042/R001.19.0.1812568, 232043/R001.19.0.1812568, 232044/R001.19.0.1812568, 232045/R001.19.0.1812568, 232046/R001.19.0.1812568, 232047/R001.19.0.1812568, 232048/R001.19.1.1929746, 232049/R001.19.1.1929746, 232050/R001.19.1.1929746, 232051/R001.19.1.1929746, 202023/R001.19.1.1929746, 202024/R001.19.1.1929746, 232052/R001.19.1.1929746, 232053/R001.19.1.1929746
    FDA recall number
    Z-3283-2026
    Quantity
    34
  • Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama 35 with software

    Lot or code
    UDI-DI: 06971576832170. Serial Number: 11MI410003/R001.19.4.1909524, 11MI410004/R001.19.4.1909524, 11MI410005/R001.19.4.1909524, 20MI410001/R001.19.4.1909524, 20MI410002/R001.19.4.1909524, 20MI410003/R001.21.0.1993334, 20MI410004/R001.19.4.1909524
    FDA recall number
    Z-3284-2026
    Quantity
    7
  • Positron Emission Tomography and Computed Tomography System, Model: uMI Panorama GS with software

    Lot or code
    UDI-DI: 06971576832187. Serial Number: 11MIP00007/R001.17.5.1785268, 11MIP00009/R001.18.2.1967222, 11MIP00010/R001.18.2.1967222
    FDA recall number
    Z-3285-2026
    Quantity
    3
  • Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo with software

    Lot or code
    UDI-DI: 06971576832231. Serial Number/Software: 11MI500004/R002.4.0.1872339, 11MI500005/R002.4.0.1872339, 11MI500006/R002.4.0.1872339
    FDA recall number
    Z-3286-2026
    Quantity
    3
  • Positron Emission Tomography and Computed Tomography System, Model: uMI Panvivo S with software

    Lot or code
    UDI-DI: 06971576832255. Serial Number/Software: 11MI510005/R003.2.0.1872339, 20MI510001/R003.2.0.1872339, 20MI510002/R003.2.0.1872339, 20MI510003/R003.2.0.1872339, 20MI510004/R003.2.0.1872339, 20MI510005/R003.2.0.1872339
    FDA recall number
    Z-3287-2026
    Quantity
    6
  • Positron Emission Tomography and Computed Tomography System, Model: uMI 780

    Lot or code
    UDI-DI: 06971576832033. Serial Number/Software: 300098/R001. 19.0.1812568
    FDA recall number
    Z-3288-2026
    Quantity
    1

Distribution (FDA): US Nationwide distribution in the states of CA, PR, UT, WV, ND, NE, NC, TX, ID, LA, MT, FL, CO, IN, KY, AL, OH, MI, IA, MO, NY, NV, TN, WV, SC, WI, IL, MS, MA, NJ.

Source

Issued by
U.S. Food and Drug Administration
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2026, Week 2026-W41. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99731 on October 7, 2026.