Why it was recalled
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
1 more codes are in the record below.
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
- Lot or code
- UDI: (01)05391523442800/Lot # 65705261
- FDA recall number
- Z-0205-2027
- Quantity
- 832 units
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
- Lot or code
- UDI:(01)05391523442817/Lot # 65705262
- FDA recall number
- Z-0206-2027
- Quantity
- 123 units
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
- Lot or code
- UDI:(01)05391523442824/-Lot # 65705263
- FDA recall number
- Z-0207-2027
- Quantity
- 832 units