What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
- Lot or code
- All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI(s): (01)00884838102514 (01)00884838107540 (01)00884838114159 (01)00884838124912 (01)00884838128125
- FDA recall number
- Z-3249-2026
- Quantity
- 219 units
Distribution (FDA): Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Philips Ultrasound, LLC (other notices of this firm)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.