Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99704: Steris Corporation: VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006 and 5 more

FDA recall 99704 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

5 more codes are in the record below.

  • VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

    Lot or code
    UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027
    FDA recall number
    Z-0070-2027
    Quantity
    266
  • VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB007

    Lot or code
    UDI: 10724995220331 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
    FDA recall number
    Z-0071-2027
    Quantity
    409
  • VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

    Lot or code
    UDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027
    FDA recall number
    Z-0072-2027
    Quantity
    1185
  • VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061

    Lot or code
    UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027
    FDA recall number
    Z-0073-2027
    Quantity
    2205
  • VERIFY Biological Steam Test Pack Model/Catalog Number: S3065

    Lot or code
    UDI: 10724995220300 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
    FDA recall number
    Z-0074-2027
    Quantity
    217
  • VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

    Lot or code
    UDI:10724995023864 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027
    FDA recall number
    Z-0075-2027
    Quantity
    233

Distribution (FDA): Worldwide - US Nationwide distribution.

Source

Issued by
U.S. Food and Drug Administration
Firm
Steris Corporation (other notices of this firm)
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Steris Corporation, Month: October 2026, Week 2026-W42. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99704 on October 14, 2026.

Other notices of Steris Corporation