What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
4 more codes are in the record below.
Convenience kits ANESTHESIA PAIN PK LEAGUE DYNJ67925A
- Lot or code
- UDI-DI 10195327506506 Lot 25BBK494
- FDA recall number
- Z-3207-2026
- Quantity
- 341
Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F
- Lot or code
- DYNJ58421A UDI-DI 10193489970487 Lot 25HBQ049 DYNJ40734F UDI-DI 10198459162084 Lots 25GBO299 25IBS607
- FDA recall number
- Z-3208-2026
- Quantity
- 459
Convenience kits MYRINGOTOMY PACK-LF DYNJ59097A
- Lot or code
- UDI-DI 10198459046551 Lot 25FBU754
- FDA recall number
- Z-3209-2026
- Quantity
- 60
Convenience kits CATARACT PACK DYNJ59881A
- Lot or code
- UDI-DI 10198459020612 Lot 25GBP246
- FDA recall number
- Z-3210-2026
- Quantity
- 336
Distribution (FDA): US Nationwide distribution in states of Arkansas, Florida, North Carolina, Maryland, and Texas.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medline Industries, LP (other notices of this firm)
- Title