What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.
FDA classification: Class I. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.
- Lot or code
- Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)
- FDA recall number
- Z-3090-2026
- Quantity
- 127,416
Distribution (FDA): Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Given Imaging Ltd.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.