Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99467: Medos International Sarl: Velys Enabling Tech Robotic-Assist Station REF: 451611001

FDA recall 99467 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

23 more codes are in the record below.

  • Velys Enabling Tech Robotic-Assist Station REF: 451611001

    Lot or code
    UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901
    FDA recall number
    Z-3088-2026
    Quantity
    29 stations

Distribution (FDA): US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.

Source

Issued by
U.S. Food and Drug Administration
Firm
Medos International Sarl
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: September 2026, Week 2026-W37. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99467 on September 9, 2026.