Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99459: Biofire Defense: FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code:

FDA recall 99459 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

  • FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).

    Lot or code
    Part Number: NGDS-ASY-0007 UDI-DI code: (01)00851458005136 Lot Numbers: 311825D 312025D 313526D 322925D
    FDA recall number
    Z-2877-2026
    Quantity
    122 kits (3,660 pouches)

Distribution (FDA): U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.

Source

Issued by
U.S. Food and Drug Administration
Firm
Biofire Defense
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Firm: Biofire Defense, Month: August 2026, Week 2026-W33. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99459 on August 12, 2026.

Other notices of Biofire Defense