Skip to content
FDAMedical devicesClass IIOngoing

FDA recall 99409: Alere San Diego, Inc.: McKesson CONSULT hCG Urine Cassette and 3 more

FDA recall 99409 ·

Build snapshot. Collected , once, when this site was built. It is not a live feed.

Check my lot or code

Is yours in this recall?

Type the lot, batch, date code, model number or UPC printed on the product. It is compared here, in your browser, with the codes this notice lists. It stays in this page: it is not added to the address.

General information on recalls from the agency: U.S. Food and Drug Administration (opens in a new tab).

What the agency says to do

FDA’s record gives no instruction for consumers.

Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)

Note from Recalls Watch, not agency text

Why it was recalled

hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Text from FDA, unedited.

FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)

The recalled product

Codes in the record. Click one to copy it.

77 more codes are in the record below.

  • McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/Urine¿hCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;

    Lot or code
    All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 5001/ 20612479202744/1155958203, 12/15; 92210/10815845020130/1156015803, 07/16; 23900530/00613647000571/1155810105, 2018-01-15; 23900531/00613647000533/1155810105, 2018-01-15; 5002/20612479202768/1155958305, 12/15; 92215/10815845020123/1156015903, 07/16; B-1077-23/20885380020268/1155904805, 9/2014; MPH22025HCG/10884389191880/1156093402, 04-01-2015
    FDA recall number
    Z-3271-2026
    Quantity
    50,265,835
  • Fisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+ hCG Combo test in Serum or Urine, Model/ Part: FHC-202/9004075; Alere hCG Combo Cassette (25mIU/mL) in Urine (CLIA Waived) or Serum, Model/ Part: FHC-202/92225; Pro Advantage By NDC Urine/Serum Pregnancy Cassette Device, Model/ Part: FHC-202/P080021

    Lot or code
    All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900529/00613647000540/1155810906, 2018-01-15; 23900528/00613647000564/1155810906, 2018-01-15; 43025/15099590776886/May 2015; 9004075/10304040024123/1155875804, 2014-06-09; 92225/10815845020116/1156015603, 07/16; P080021/M220P0800213/40001856, 2023-03-30
    FDA recall number
    Z-3272-2026
    Quantity
    5,331,585
  • Fisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine Test Dipstick CLIA Waived, Model/ Part:FHC-101/5000; Henry Schein One Step+ hCG Urine Strip Test, Model/ Part: FHC-101/9004072; Alere hCG Dipstick (25mIU/mL) in Urine, Model/ Part: FHC-101/92211; Cardinal Health hCG Dipstick Rapid Test CLIA Waived, Model/ Part: FHC-101/B-1077-21; Medline hCG Pregnancy Test Strip, Model/ Part: FHC-101/MPH11025HCG; Pro Advantage by NDC Urine Pregnancy Test Strip, Model/ Part: FHC-101/P080023;

    Lot or code
    All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900527/00613647000588/1155811305, 2018-01-15; 5000/20612479202751/40008059, V05; 9004072/10304040024093/1155800504, 2014-06-09; 92211/10815845020093/40008089, 05/25; B-1077-21/20885380020244/40008054, 04/2025; MPH11025HCG/10884389191873/40006717, 6/2025; P080023/M220P0800233/40008087, 2025-05-09
    FDA recall number
    Z-3273-2026
    Quantity
    13,256,075
  • Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;

    Lot or code
    All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900526/00613647000526/1155811105, 2018-01-15; 9004074/10304040024116/1155875704, 2014-06-09; 9004073/10304040024109/1155875704, 2014-06-09; 92217/10815845020109/1156015703, 07/16; B1077-22/20885380020251/1155904704, 9/2014; MPH12025HCG/10888277191877/1156093302, 2015-03-20; P080063/M220P0800633/40001853, 2023-03-30
    FDA recall number
    Z-3274-2026
    Quantity
    18,139,960

Distribution (FDA): Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.

Source

Issued by
U.S. Food and Drug Administration
Firm
Alere San Diego, Inc.
Title
Composed by Recalls Watch from fields of the record; FDA records have no title.

Check all notices that name this firm in the directory of recalls by brand and manufacturer. This notice is in these lists: Month: October 2026, Week 2026-W41. Browse more medical devices recalls, or check another notice by its number in the recall number lookup.

About this recall

What is the hazard?

FDA states: “hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.”

What should I do if I have this product?

FDA’s record gives no instruction for consumers. The official report is linked on this page.

How do I know if my product is included?

Compare the lot or code printed on your product with the codes FDA lists on this notice, product by product.

How does the code check on this page work?

You type the lot, batch, date code, model number or UPC from your package and the page compares it, in your browser, with the codes FDA lists on this notice. It reports a match, a possible match (your code is part of a longer listed code) or no match. It only compares text: a notice that describes products by dates, sizes or “all lots” cannot be compared, so “no match” does not clear a product. The comparison runs in your browser and the page does not send the code anywhere, and the code stays out of the page address. Only if you press “Copy link to this result” is a link made that carries it, and whoever opens that link requests that address from this site.

When was this recall published?

FDA published recall 99409 on October 7, 2026.