What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
- Lot or code
- UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
- FDA recall number
- Z-2775-2026
- Quantity
- 77
Distribution (FDA): US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- KENT IMAGING, INC.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.