What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
The affected Endo-Model replacement set for a Rotating Hinge tibial plateau incorrectly contains a Pure Hinge plateau. The incorrect plateau could lead to a prolonged or modified surgery as the plateau cannot be inserted and an alternative product would be needed. If the incorrect plateau were to be implanted, this is associated with increased risk of wear, which could result in a reduced standing time of the implant.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Endo-Model, Replacement Set, Tibial Plateau, right/left, L, UHMWPE, CoCrMo, Rotating Hinge Item Number 15-0027/16
- Lot or code
- UDI-DI 04026575432134 Lot for Implant REF 15-0027/16: 2604069
- FDA recall number
- Z-2934-2026
- Quantity
- 1 US; 9 OUS
Distribution (FDA): Domestic US distribution to Indiana. OUS distribution to Belarus and Germany.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.