What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Power Over Fiber Base Module
- Lot or code
- Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731
- FDA recall number
- Z-2745-2026
- Quantity
- 101
Distribution (FDA): U.S. Nationwide Distribution and International Distribution
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Terex Utilities
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.