What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
Ventriculostomy Kit, REF: 46154
- Lot or code
- UDI-DI: 00673978947396, Lot: 2025-08779
- FDA recall number
- Z-2768-2026
- Quantity
- 167
Distribution (FDA): Worldwide - US Nationwide distribution in the states of UT, AL, CA, NC, GA, AR, LA, KY, LA, KY, IL, OK and the country of Costa Rica.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- Medtronic Neurosurgery
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.