Why it was recalled
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
- Lot or code
- UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3
- FDA recall number
- Z-2938-2026
- Quantity
- 1126 units
Distribution (FDA): US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.