What the agency says to do
FDA’s record gives no instruction for consumers.
Status: Ongoing. What the FDA’s recall statuses mean (opens in a new tab)
Note from Recalls Watch, not agency text
- In an emergency call 911. For a possible poisoning call Poison Help at 1-800-222-1222. Poison Help (HRSA) (opens in a new tab)
- For refunds, repairs or replacements, follow the agency notice. Recalls.gov (opens in a new tab)
Why it was recalled
Demineralized bone matrix (DBM) in resorbable mesh indicated for use with autograft as a bone graft extender in the posterolateral spine and pelvis, were filled with approx. 4.5 cc DBM (fill for 45 ¿ 11 mm configuration) instead of approx. 7.5 cc DBM (fill for 45 ¿ 17.5 mm configuration), which may lead to delayed fusion, incomplete fusion, pseudarthrosis, or lack of bone ingrowth.
FDA classification: Class II. What the FDA’s classes mean (opens in a new tab)
The recalled product
Codes in the record. Click one to copy it.
OsteoBallast, 45 x 17.5 mm (SeaSpine Brand), REF: 56800095. Ballast, 45 x 17.5 mm (IsoTis Brand), REF: 02-8000-095
- Lot or code
- REF/UDI-DI/Lots: 56800095/10889981149352/181559, 181550 02-8000-095/10889981149314/181559, 181550
- FDA recall number
- Z-2820-2026
- Quantity
- 98
Distribution (FDA): US Nationwide distribution in the states of AK, NJ, TN, AR, CA, WA, OH.
Source
- Issued by
- U.S. Food and Drug Administration
- Firm
- IsoTis OrthoBiologics, Inc.
- Title
- Composed by Recalls Watch from fields of the record; FDA records have no title.